Korean J Lab Med.  2007 Dec;27(6):394-399. 10.3343/kjlm.2007.27.6.394.

New Leukocyte Counting Method of Cerebrospinal Fluid: Using Transparent Ruler Tape

Affiliations
  • 1Department of Laboratory Medicine, College of Medicine, Kangwon National University, Chuncheon, Korea. bloodmd@kangwon.ac.kr

Abstract

BACKGROUND: To enumerate leukocyte count in cerebrospinal fluid (CSF) is important for diagnosing bacterial meningitis. Using automated hematology analyzer for enumeration of leukocyte in CSF is below the sensitivity, so microscopic hemocytometric method is standard method. But this requires sufficient practical experience and has limitation of accuracy and stability. So we developed new manual method and evaluated it. METHODS: We designed new method using transparent ruler tape. We performed correlation, accuracy and precision test by counting leukocyte in diluted EDTA blood with three methods: new method, Neubauer and Nageotte hemocytometry. Twenty two CSF were used for stability test, which determines leukocyte count according to time (within one hour and after 2, 4 and 12 hr), by new method and Neubauer hemocytometry at room temperature. RESULTS: There was no clinical significant difference between three methods in correlation test, whereas Neubauer and Nageotte hemocytometry showed a bias to underestimation relative to the results obtained with new method in case with low leukocyte count. The new method showed the lowest CV and most accurate result. In stability test, leukocyte counts decreased being 44.4%, 72.1% of initial values after 2 hr, 14.8%, 31.1%, after 4 hr and 4.2%, 8.7%, after 12 hr, by Nageotte hemocytometry and new method, respectively. CONCLUSIONS: The new method we devised is simple, easy and applicable to use in a laboratory and offers advantages of improved precision and stability. It may be sufficient for replacing standard methods for leukocyte counting in CSF.

Keyword

CSF; Leukocyte count; Transparent ruler tape; New manual method

MeSH Terms

Cerebrospinal Fluid/*cytology
Female
Humans
Leukocyte Count/instrumentation/*methods
Male
Regression Analysis
Reproducibility of Results
Sensitivity and Specificity
Time Factors

Figure

  • Fig. 1. Slide with transparent ruler tape for cell count (A), visual fields of microscope ×40 (B), ×100 (C), ×200 (D), ×400 (E).

  • Fig. 2. Schematic figure of new counting method. The arrow indicates direction in microscopic examination.

  • Fig. 3. Microscopic finding of CSF slide with transparent ruler tape (×100).

  • Fig. 4. Linearity of leukocyte counting with three method; Neubauer hemacytometer (A), Nageotte chamber (B), new counting method (C) using diluents of EDTA blood samples. Each point represent an individual observation (N=10).

  • Fig. 5. Leukocyte count (/μL) change with Nageotte chamber (A) and new counting method (B) of each samples (6 cases of initial WBC count >50/μL) according to time elapse.


Reference

1.Hoen B., Viel JF., Paquot C., Gerard A., Canton P. Multivariate approach to differential diagnosis of acute meningitis. Eur J Clin Microbiol Infect Dis. 1995. 14:267–74.
Article
2.Muller TH., Doscher A., Schunter F., Scott CS. Manual and automated methods for the determination of leukocyte counts at extreme low levels: comparative evaluation of the Nageotte chamber and the Abbott Cell Dyn 3500 analyser. Transfus Sci. 1997. 18:505–15.
3.Chow G., Schmidley JW. Lysis of erythrocytes and leukocytes in traumatic lumbar punctures. Arch Neurol. 1984. 41:1084–5.
Article
4.Veerman AJ., Huismans L., van Zantwijk I. Storage of cerebrospinal fluid samples at room temperature. Acta Cytol. 1985. 29:188–9.
5.Steele RW., Marmer DJ., O'Brien MD., Tyson ST., Steele CR. Leukocyte survival in cerebrospinal fluid. J Clin Microbiol. 1986. 23:965–6.
Article
6.Ziebig R., Lun A., Sinha P. Leukocyte counts in cerebrospinal fluid with the automated hematology analyzer CellDyn 3500 and the urine flow cytometer UF-100. Clin Chem. 2000. 46:242–7.
Article
7.Andrews J., Setran E., McDonnel L., Kussick S., Wood BL., Sabath DE. An evaluation of the cell-dyn 3200 for counting cells in cerebrospinal and other body fluids. Lab Hematol. 2005. 11:98–106.
Article
8.Van Acker JT., Delanghe JR., Langlois MR., Taes YE., De Buyzere ML., Verstraete AG. Automated flow cytometric analysis of cerebrospinal fluid. Clin Chem. 2001. 47:556–60.
Article
9.Brecher ME., Harbaugh CA., Pineda AA. Accurate counting of low numbers of leukocytes. Use of flow cytometry and a manual low-count chamber. Am J Clin Pathol. 1992. 97:872–5.
Article
10.Dzik S., Moroff G., Dumont L. A multicenter study evaluating three methods for counting residual WBCs in WBC-reduced blood components: Nageotte hemocytometry, flow cytometry, and microfluorometry. Transfusion. 2000. 40:513–20.
Article
11.Borzini P., Dumont LJ. Microdroplet fluorochromatic assay for the enumeration of white cells (WBCs) in WBC-reduced blood components: validation and application for evaluating newly developed WBC-reduction filters. Transfusion. 1997. 37:601–6.
Article
12.Seghatchian J., Krailadsiri P., Scott CS. Counting of residual WBCs in WBC-reduced blood components: a multicenter evaluation of a microvolume fluorimeter by the United Kingdom National Blood Service. Transfusion. 2001. 41:93–101.
Article
13.Cassens U., Greve B., Tapernon K., Nave B., Severin E., Sibrowski W, et al. A novel true volumetric method for the determination of residual leucocytes in blood components. Vox Sang. 2002. 82:198–206.
Article
14.Watson MA., Scott MG. Clinical utility of biochemical analysis of cerebrospinal fluid. Clin Chem. 1995. 41:343–60.
Article
15.Spanos A., Harrell FE Jr., Durack DT. Differential diagnosis of acute meningitis. An analysis of the predictive value of initial observations. JAMA. 1989. 262:2700–7.
Article
16.Nelson RP Jr. Bacterial meningitis and inflammation. Curr Opin Neurol. 2006. 19:369–73.
Article
17.Fishbein DB., Palmer DL., Porter KM., Reed WP. Bacterial meningitis in the absence of CSF pleocytosis. Arch Intern Med. 1981. 141:1369–72.
Article
18.Lindquist L., Linne T., Hansson LO., Kalin M., Axelsson G. Value of cerebrospinal fluid analysis in the differential diagnosis of meningitis: a study in 710 patients with suspected central nervous system infection. Eur J Clin Microbiol Infect Dis. 1988. 7:374–80.
Article
19.Nye FJ. The value of initial laboratory investigations in the management of meningitis. J Infect. 1983. 7:31–8.
Article
20.Aulesa C., Mainar I., Prieto M., Cobos N., Galimany R. Use of the Advia 120 hematology analyzer in the differential cytologic analysis of biological fluids (cerebrospinal, peritoneal, pleural, pericardial, synovial, and others). Lab Hematol. 2003. 9:214–24.
21.Barnes PW., Eby CS., Shimer G. An evaluation of the utility of performing body fluid counts on the coulter LH 750. Lab Hematol. 2004. 10:127–31.
Article
22.de Jonge R., Brouwer R., van Rijn M., van Acker BA., Otten HJ., Lindemans J. Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100. Clin Chem Lab Med. 2006. 44:1367–71.
Article
Full Text Links
  • KJLM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr