Korean J Oral Maxillofac Radiol.  2001 Mar;31(1):1-7.

Comparison of personal computer with CT workstation in the evaluation of 3-dimensional CT image of the skull

Affiliations
  • 1Department of Dental Radiology, College of Dentistry, Yonsei University, Korea. kdkim@yonsei.ac.kr

Abstract

PURPOSE: To evaluate the usefulness of the reconstructed 3-dimensional image on the personal computer in comparison with that of the CT workstation by quantitative comparison and analysis.
MATERIALS AND METHODS
The spiral CT data obtained from 27 persons were transferred from the CT workstation to a personal computer, and they were reconstructed as 3-dimensional image on the personal computer using V-works 2.0(TM). One observer obtained the 14 measurements on the reconstructed 3-dimensional image on both the CT workstation and the personal computer. Paired t test was used to evaluate the intraobserver difference and the mean value of the each measurement on the CT workstation and the personal computer. Pearson correlation analysis and % incongruence were also performed.
RESULTS
I-Gn, N-Gn, N-A, N-Ns, B-A, and G-Op did not show any statistically significant difference (p>0.05), B-O, B-N, Eu-Eu, Zy-Zy, Biw, D-D, Orbrd R, and L had statistically significant difference (p<0.05), but the mean values of the differences of all measurements were below 2 mm, except for D-D. The value of correlation coefficient gamma was greater than 0.95 at I-Gn, N-Gn, N-A, N-Ns, B-A, B-N, G-Op, Eu-Eu, Zy-Zy, and Biw, and it was 0.75 at B-O, 0.78 at D-D, and 0.82 at both Orbrd R and L. The % incongruence was below 4% at I-Gn, N-Gn, N-A, N-Ns, B-A, B-N, G-Op, Eu-Eu, Zy-Zy, and Biw, and 7.18%, 10.78%, 4.97%, 5.89% at B-O, D-D, Orbrd R and L respectively.
CONCLUSION
It can be considered that the utilization of the personal computer has great usefulness in reconstruction of the 3-dimensional image when it comes to the economics, accessibility and convenience, except for thin bones and the landmarks which are difficult to be located.

Keyword

CT workstation; personal computer; reconstructed 3-dimensional CT image

MeSH Terms

Humans
Microcomputers*
Skull*
Tomography, Spiral Computed