J Korean Soc Ther Radiol Oncol.  2010 Mar;28(1):50-56. 10.3857/jkstro.2010.28.1.50.

External Auditing on Absorbed Dose Using a Solid Water Phantom for Domestic Radiotherapy Facilities

Affiliations
  • 1Department of Radiation Applying Life Science, Seoul National University Graduate School, Seoul, Korea. sye@snu.ac.kr
  • 2Institute of Radiation Medicine, Medical Research Center, Seoul National University College of Medicine, Seoul, Korea.
  • 3Radiation Research, Korean Institute of Nuclear Safety, Seoul, Korea.
  • 4Korea Food and Drug Administration, Seoul, Korea.
  • 5Department of Radiation Oncology, Seoul National University College of Medicine, Seoul, Korea.

Abstract

PURPOSE
We report the results of an external audit on the absorbed dose of radiotherapy beams independently performed by third parties. For this effort, we developed a method to measure the absorbed dose to water in an easy and convenient setup of solid water phantom.
MATERIALS AND METHODS
In 2008, 12 radiotherapy centers voluntarily participated in the external auditing program and 47 beams of X-ray and electron were independently calibrated by the third party's American Association of Physicists in Medicine (AAPM) task group (TG)-51 protocol. Even though the AAPM TG-51 protocol recommended the use of water, water as a phantom has a few disadvantages, especially in a busy clinic. Instead, we used solid water phantom due to its reproducibility and convenience in terms of setup and transport. Dose conversion factors between solid water and water were determined for photon and electron beams of various energies by using a scaling method and experimental measurements.
RESULTS
Most of the beams (74%) were within +/-2% of the deviation from the third party's protocol. However, two of 20 X-ray beams and three of 27 electron beams were out of the tolerance (+/-3%), including two beams with a >10% deviation. X-ray beams of higher than 6 MV had no conversion factors, while a 6 MV absorbed dose to a solid water phantom was 0.4% less than the dose to water. The electron dose conversion factors between the solid water phantom and water were determined: The higher the electron energy, the less is the conversion factor. The total uncertainty of the TG-51 protocol measurement using a solid water phantom was determined to be +/-1.5%.
CONCLUSION
The developed method was successfully applied for the external auditing program, which could be evolved into a credential program of multi-institutional clinical trials. This dosimetry saved time for measuring doses as well as decreased the uncertainty of measurement possibly resulting from the reference setup in water.

Keyword

Quality assurance; External auditing; Solid water phantom; Dosimetry