J Korean Assoc Maxillofac Plast Reconstr Surg.  2011 May;33(3):233-240.

Measurement and Algorithm Calculation of Maxillary Positioning Change by Use of an Optoelectronic Tracking System Marker in Orthognathic Surgery

Affiliations
  • 1Department of Dentistry, School of Dentistry, Seoul National University, Korea. smin5@snu.ac.kr
  • 2Department of Oral and Maxillofacial Surgery, School of Dentistry, Seoul National University, Korea.
  • 3Dental Research Institute, Seoul National University, Korea.

Abstract

PURPOSE
To apply a computer assisted navigation system to orthognathic surgery, a simple and efficient measuring algorithm calculation based on affine transformation was designed. A method of improving accuracy and reducing errors in orthognathic surgery by use of an optical tracking camera was studied.
METHODS
A total of 5 points on one surgical splint were measured and tracked by the Polaris Vicra(R) (Northern Digital Inc Co., Ontario, Canada) optical tracking system in two cases. The first case was to apply the transformation matrix at pre- and postoperative situations, and the second case was to apply an affine transformation only after the postoperative situation. In each situation, the predictive measuring value was changed to the final measuring value via an affine transformation algorithm and the expected coordinates calculated from the model were compared with those of the patient in the operation room.
RESULTS
The mean measuring error was 1.027+/-0.587 using the affine transformation at pre- and postoperative situations and the average value after the postoperative situation was 0.928+/-0.549. The farther a coordinate region was from the reference coordinates which constitutes the transform matrixes, the bigger the measuring error was found which was calculated from an affine transformation algorithm.
CONCLUSION
Most difference errors were brought from mainly measuring process and lack of reproducibility, the affine transformation algorithm formula from postoperative measuring values by using of optic tracking system between those of model surgery and those of patient surgery can be selected as minimizing the difference error. To reduce coordinate calculation errors, minimum transformation matrices must be used and reference points which determine an affine transformation must be close to the area where coordinates are measured and calculated, as well as the reference points need to be scattered.

Keyword

Affine transformation; Coordinate measurement; Navigation surgery; Orthognathic surgery; Point maker