Journal Browser Advanced Search Help
Journal Browser Advanced search HELP
Hip Pelvis. 2017 Sep;29(3):199-203. English. Case Report. https://doi.org/10.5371/hp.2017.29.3.199
Santoso A , Choi IS , Park KS , Yoon TR .
Department of Orthopaedic and Traumatology, Faculty of Medicine, Sebelas Maret University, Prof. Dr. R. Soeharso Orthopaedic Hospital, Solo, Indonesia.
Department of Orthopedic Surgery, Center for Joint Disease, Chonnam National University Hwasun Hospital, Hwasun, Korea. chiasma@hanmail.net
Abstract

Proximal femoral cephalomedullary nailing is one of the treatments of choice for intertrochanteric fracture. Zimmer Natural NailĀ® (ZNN; Zimmer) is one of the alternatives. We report two cases of broken tail portion of the ZNN lag screw during the removal procedure. This may be the first reported cases in scientific literature with this pattern of failure. We report these cases to highlight one of the possible problems that could occur during removal of this implant which can possibly prolong the operation time significantly if the required device is not prepared.

Copyright © 2019. Korean Association of Medical Journal Editors.