Journal Browser Advanced Search Help
Journal Browser Advanced search HELP
J Korean Soc Ther Radiol. 1987 Dec;5(2):137-140. English. Original Article.
Huh SJ , Ha SW , Choi KY .
Department of Radiation Therapy, Soon Chun Hyang University Hospital, Seoul, Korea.
Department of Radiation Therapy, College of Medicine, Seoul National Univerisy, Seoul, Korea.
Abstract

In brachytherapy of uterine cervical cancer using a high dose rate remote afterloading system, it is of prime importance to deliver a accurate dose in each fractionated treatment by minimizing the difference between the pre-treatment planned and post-treatment calculated doses. The post-treatment calculated point A dose was not much different from the pretreatment planned dose (500 cGy). The average+/-standard deviation was 500+/-18 cGy and 84 percent of 82 intracavitary radiotherapy was within the range of 500+/-25 cGy.

Copyright © 2019. Korean Association of Medical Journal Editors.