Korean Diabetes J. 2010 Dec;34(6):350-358. English. Original Article. https://doi.org/10.4093/kdj.2010.34.6.350
Baek YH , Jin HY , Lee KA , Kang SM , Kim WJ , Kim MG , Park JH , Chae SW , Baek HS , Park TS .
Division of Endocrinology and Metabolism, Department of Internal Medicine, Chonbuk National University Hospital, Chonbuk National University Medical School, Jeonju, Korea. pts@jbnu.ac.kr
Department of Pharmacology, Chonbuk National University Medical School, Jeonju, Korea.
Clinical Trial Center, Chonbuk National University Hospital, Chonbuk National University Medical School, Jeonju, Korea.
Abstract

BACKGROUND: Clinical experience with the continuous glucose monitoring systems (CGMS) is limited in Korea. The objective of this study is to evaluate the accuracy of the CGMS and the correlation between interstitial fluid and venous plasma glucose level in Korean healthy male subjects. METHODS: Thirty-two subjects were served with glucose solution contained same amount of test food's carbohydrate and test foods after separate overnight fasts. CGMS was performed over 3 days during hopitalization for each subjects. Venous plasma glucose measurements were carried out during 4 hours (0, 0.25, 0.5, 0.75, 1, 2, 4 hours) just before and after glucose solution and test food load. The performance of the CGMS was evaluated by comparing its readings to those obtained at the same time by the hexokinase method using the auto biochemistry machine (Hitachi 7600-110). Also, correlations between glucose recorded with CGMS and venous plasma glucose value were examined. RESULTS: CGMS slightly underestimated the glucose value as compared with the venous plasma glucose level (16.3 +/- 22.2 mg/dL). Correlation between CGMS and venous plasma glucose values throughout sensor lifetime is 0.73 (regression analysis: slope = 1.08, intercept = 8.38 mg/dL). Sensor sensitivity can deteriorate over time, with correlations between venous blood glucose and CGMS values dropping from 0.77 during 1st day to 0.65 during 2nd and 3rd day. CONCLUSION: The accuracy of data provided by CGMS may be less than expected. CGMS sensor sensitivity is decreased with the passage of time. But, from this study, CGMS can be used for glucose variability tendency monitoring conveniently to the Korean.

Copyright © 2019. Korean Association of Medical Journal Editors.