Yonsei Med J.  1987 Mar;28(1):71-74. 10.3349/ymj.1987.28.1.71.

A Failure of the Locking System of the Acetabular Assembly in a Total Hip Replacement

Affiliations
  • 1Department of Orthopedic Surgery, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea.

Abstract

In this study, the author presents the case of one patient who had a Harris-Galante total hip replacement for pigmented villonodular synovitis of the left hip joint, the locking system failed and separation of the polyethylene liner from the acetabular metal shell occurred while the closed reduction of the posterior dislocation of the femoral prosthesis was performed. This study indicates that the design of the acetabular assembly should be modified. The gap between the rims of metal shell and polyethylene liner should be reduced or eliminated.

Keyword

Total hip replacement; dislocation of femoral prosthesis; separation of polyethylene liner

MeSH Terms

Acetabulum*/radiography
Hip Prosthesis*
Human
Male
Middle Age
Postoperative Complications/radiography*
Prosthesis Failure
Synovitis/surgery*
Synovitis, Pigmented Villonodular/surgery*