J Korean Med Sci.  2001 Dec;16(6):702-706. 10.3346/jkms.2001.16.6.702.

The Flow Rate of the Elastomeric Balloon Infusor is Influenced by the Internal Pressure of the Infusor

Affiliations
  • 1Department of Anesthesiology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. jakim01@samsung.co.kr
  • 2Department of Anesthesiology, College of Medicine, Seoul National University, Seoul, Korea.

Abstract

The purpose of this study was to evaluate the flow rate of elastomeric balloon infusor composed of rubber or silicone materials. Two models were studied: the Baxter Twoday Infusor and Advance Silicone Infusor. Each infusion device has a preset flow rate of 2 mL/hr. Fifteen units of each device were filled with 100 mL of normal saline. The flow rate was measured gravimetically using an electronic balance. The internal pressure of the balloon infusor was measured with a Digital pressure meter via the pressure transducer. The monitored internal pressure of the two kinds of infusor was not maintained uniformly during the entire delivery period, which was divided into 3 phases: the phase of decreasing, maintained, and increasing pressure during the delivery of drug. Both devices initially infused at a relatively high rate, followed by a somewhat steady flow rate. The flow rate distinctly increased in the small residual volume. The flow rate of the balloon infusor used in this study was not sustained uniformly during the entire delivery period and was in proportion to the internal pressure of the infusor regardless of the materials.

Keyword

Elastomers; Infusion Pumps; Flow rate

MeSH Terms

Analgesia, Patient-Controlled/*instrumentation
Elasticity
Equipment Design
Human
*Infusion Pumps
Pain, Postoperative/*therapy
Pressure
Full Text Links
  • JKMS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr