Korean J Ophthalmol.  2024 Dec;38(6):496-506. 10.3341/kjo.2024.0115.

Repeatability of Corneal Epithelial Thickness Map Using Anterior-Segment Optical Coherence Tomography in Normal and Corneal Disease Patients

Affiliations
  • 1Department of Ophthalmology, Ewha Womans University Mokdong Hospital, Ewha Womans University College of Medicine, Seoul, Korea

Abstract

Purpose
To evaluate the repeatability of corneal epithelial thickness (CET) measurements in normal eyes and eyes diagnosed with corneal disease using the epithelial thickness map (ETM) of anterior-segment optical coherence tomography (OCT).
Methods
In this retrospective study, patients with three OCT scans using the ETM mode of Cirrus OCT between October 2021 and January 2024 were reviewed. Two groups of subjects were included: (1) normal subjects with no history of ophthalmic surgery, corneal diseases, and topical antiglaucoma medication uses; and (2) subjects with corneal diseases including dry eye syndrome, recurrent corneal erosion, pterygium, and others. A total of 57 eyes of 57 normal subjects and 106 eyes of 76 patients with corneal disease were included. ETM was analyzed in 25 zones (one zone within 0–2 mm diameter, eight zones within 2–5 mm diameter, eight zones within 5–7 mm diameter, and eight zones within 7–9 mm diameter). Repeatability was evaluated by calculating intraclass correlation coefficient (ICC), coefficient of variation (CoV), within-subject standard deviation (Sw), and Bland-Altman plot.
Results
Among a total of 25 sectors, the normal eyes showed high repeatability (ICC, >0.75; CoV, 2.160%–5.292%; Sw, 0.760– 1.653 μm) in 23 sectors, and corneal diseases patients also showed high repeatability (ICC, >0.75; CoV, 4.167%–9.606%; Sw, 1.298–3.340 μm) in 22 sectors. However, the wide range of 95% limit of agreement width of Bland-Altman plot presented in corneal disease group and some peripheral zones in normal eyes indicates some variability of CET measurements.
Conclusions
Except for a few peripheral sectors, ETM of Cirrus OCT provides repeatable CET measurements in normal eyes; however in corneal disease group, repeatability was not consistently high. To measure CET accurately, performing multiple measurements is advised especially in patients with corneal disease and patients in whom peripheral CET values.

Keyword

Anterior-segment optical coherence tomography; Corneal epithelial thickness; Corneal topography
Full Text Links
  • KJO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr