Imaging Sci Dent.  2024 Sep;54(3):264-270. 10.5624/isd.20240030.

Image quality-based dose optimization in pediatric cone-beam computed tomography: A pilot methodological study

Affiliations
  • 1Department of Oral and Maxillofacial Radiology, Yonsei University College of Dentistry, Seoul, Korea
  • 2Department of Oral and Maxillofacial Radiology, Kyung Hee University Dental Hospital, Seoul, Korea

Abstract

Purpose
This study aimed to propose a methodological approach for reducing the radiation dose in pediatric conebeam computed tomography (CBCT), focusing exclusively on balancing image quality with dose optimization.
Materials and Methods
The dose-area product (DAP) for exposure was reduced using copper-plate attenuation of an X-ray source. The thickness of copper (Cu) was increased from 0 to 2.2 mm, and 10 different DAP levels were used. The QUART DVT_AP phantom and pediatric radiologic dentiform were scanned under the respective DAP levels. The contrast-to-noise ratio (CNR), image homogeneity, and modulation transfer function (MTF) were analyzed using the QUART DVT_AP phantom. An expert evaluation (overall image grade, appropriateness of field of view, artifacts, noise, and resolution) was conducted using pediatric dentiform images. The critical DAP level was determined based on phantom and dentiform analysis results.
Results
CNR and image homogeneity decreased as the DAP was reduced; however, there was an inflection point of image homogeneity at Cu 1.6 mm (DAP = 138.00 mGy·cm2 ), where the value started increasing. The MTF showed constant values as the DAP decreased. The expert evaluation of overall image grades showed “no diagnostic value” for dentiform images with Cu 1.9-2.2 mm (DAP = 78.00-103.33 mGy·cm2 ). The images with Cu 0-1.6 mm (DAP = 138.00-1697.67 mGy·cm2 ) had a “good,” “moderate,” or “poor but interpretable” grade.
Conclusion
Reducing DAP beyond a 1.6-mm Cu thickness degraded CBCT image quality. Image homogeneity and clinical image grades indicated crucial decision points for DAP reduction in pediatric CBCT scans.

Keyword

Radiation Protection; Radiation Dosage; Cone-Beam Computed Tomography; Dose Reduction