Clin Orthop Surg.  2023 Dec;15(6):960-967. 10.4055/cios22404.

Feasibility of Non-window Three-Dimensional– Printed Porous Titanium Cage in Posterior Lumbar Interbody Fusion: A Pilot Trial

Affiliations
  • 1Department of Orthopedic Surgery, Chung-Ang University Hospital, Chung-Ang University College of Medicine, Seoul, Korea
  • 2Department of Orthopedic Surgery, Sanggye Paik Hospital, Inje University College of Medicine, Seoul, Korea
  • 3Department of Orthopedic Surgery, Dongguk University Ilsan Hospital, Goyang, Korea

Abstract

Background
The commercially available design of a three-dimensional (3D)–printed titanium (3D-Ti) cage can be divided into two types according to the presence of a window: a cage with a window that allows filling of bone graft materials and a non-window cage for stand-alone use. This prospective observational case series study aimed to explore the clinical feasibility of using a nonwindow type 3D-Ti cage in cases of combined window and non-window cage implantation. Furthermore, we evaluated the bone in growth patterns of non-window cages and their correlation with published fusion grading systems.
Methods
A total of 31 consecutive patients who underwent single-level posterior lumbar interbody fusion surgery were included. Two 3D-Ti cages with different designs were inserted: a non-window cage on the left side and a window cage on the right side. Radiographic fusion was defined by the segmental angle between flexion and extension radiographs (F-E angle) and cage bridging bone (CBB) scores on computed tomography. The association between the F-E angle and osteointegration scoring system including the surface osteointegration ratio (SOR) score was analyzed.
Results
Radiographic fusion was achieved in 27 of 31 patients (87%) at 12 months postoperatively. Among the non-window cages, 23 of 31 (74.2%) had fair SOR scores, while 19 of 31 (61.3%) window cages had fair intra-cage CBB scores. The higher the SOR score was, the smaller the flexion-extension angle (SOR 0 vs. SOR 1: 6.30° ± 2.43° vs. 1.95° ± 0.99°, p < 0.001; SOR 0 vs. SOR 2: 6.03° ± 2.43° vs. 0.99°± 0.74°, p < 0.001).
Conclusions
The clinical feasibility of using a non-window 3D-Ti cage during lumbar interbody fusion might be acceptable. Furthermore, a newly suggested fusion criterion for the use of the non-window cage, the SOR score, showed a significant association with the published fusion grading systems, demonstrating its feasibility in determining interbody fusion in lumbar spinal surgery.

Keyword

Titanium; Printing; Lumbar surgery; Spinal fusion; Ossteointegration