Ultrasonography.  2023 Apr;42(2):265-274. 10.14366/usg.22135.

Intra-individual comparison of two-dimensional shear wave elastography techniques using plane wave imaging and the multi-beam technique: are they interchangeable in measuring liver fibrosis?

Affiliations
  • 1Department of Radiology, Seoul National University Hospital, Seoul, Korea
  • 2Department of Radiology, Seoul National University College of Medicine, Seoul, Korea
  • 3Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Korea

Abstract

Purpose
This study compared two different two-dimensional shear wave elastography techniques—plane wave imaging (PWI) and multi-beam (MB) imaging—from the same vendor to evaluate liver fibrosis.
Methods
In this prospective study, 42 patients with chronic liver disease who had recently undergone magnetic resonance elastography (<3 months) were enrolled, and their liver stiffness (LS) values were measured using PWI or MB. The LS values (kPa) were compared using the Wilcoxon rank-sum test. Inter-technique reproducibility and intra-observer repeatability were assessed using Bland-Altman analysis with 95% limits of agreement (LOA) and coefficients of variation (CVs). The cutoff values for predicting severe fibrosis (≥F3) were estimated using receiver operating characteristic curve (ROC) analysis, with magnetic resonance elastography as the reference standard.
Results
PWI exhibited technical failure in four patients. Therefore, 38 patients underwent both examinations. The LS values showed moderate agreement between PWI and MB (CV, 22.5%) and 95% LOA of -3.71 to 7.44 kPa. The MB technique showed good intra-observer agreement (CV, 8.1%), while PWI showed moderate agreement (CV, 11.0%). The cutoff values of PWI and MB for diagnosing ≥F3 were 12.3 kPa and 13.8 kPa, respectively, with areas under the ROC curve of 0.89 and 0.95 (sensitivity, 100% and 100%; specificity, 65.6% and 85.7%).
Conclusion
The LS values significantly differed between PWI and MB, hindering their interchangeable use in longitudinal follow-up. Considering its low technical failure rate and better repeatability, the MB technique may be preferable for evaluating liver fibrosis in chronic liver disease patients.

Keyword

Liver fibrosis; Liver stiffness measurement; 2D shear wave elastography; Plane wave imaging; Multi-beam method