Korean J Ophthalmol.  2022 Aug;36(4):306-312. 10.3341/kjo.2022.0006.

The Repeatability, Reproducibility, and Correlation of the Schirmer Test: A Comparison of Open versus Closed Eye

Affiliations
  • 1Department of Ophthalmology, Ankara Ulucanlar Eye Training and Research Hospital, Ankara, Turkey
  • 2Department of Ophthalmology, Hitit University Faculty of Medicine, Çorum, Turkey
  • 3Department of Ophthalmology, Eye Clinic, Sungurlu State Hospital, Çorum, Turkey

Abstract

Purpose
To compare the repeatability and reproducibility of the Schirmer test (ST) without anesthesia when the eyes are open (STo) and closed (STc) in previously undiagnosed patients with suggestive symptoms of dry eye.
Methods
In a comparative, observational series case study, 31 patients were included in the study. STo and STc were alternately applied for a total of six times. The ST was applied two times with the eyes open (S1) and closed (S2), respectively, by a single ophthalmologist. Then the ST was repeated four times with the eyes open (S3, S5) and closed (S4, S6), respectively, by a single nurse.
Results
S1, S3, and S5 were 23.4, 23.7, and 23.3 mm, respectively. S2, S4, and S6 were 14.7, 15.6, and 16.6 mm, respectively. STc scores were found to be statistically lower than the STo’s in general (right: t = 2.033, p = 0.048; left: t = 3.474, p = 0.004). There was no statistically significant difference in the scores of the three tests with open eyes: S1, S3, and S5 (p = 0.462). There was also no statistically significant difference in the scores of the three tests with closed eyes: S2, S4, and S6 (p = 0.05).
Conclusions
Our study suggests that although administering the ST with the patient’s eyes open produces higher readings than STc in patients with suggestive symptoms of dry eye, there was an acceptable reliability among tests performed open and closed. Moreover, intraexaminer reliability was higher than interexaminer reliability for both with the eyes open and closed.

Keyword

Dry eye syndromes; Repeatability; Reproducibility; Schirmer test
Full Text Links