Hip Pelvis.  2021 Mar;33(1):40-44. 10.5371/hp.2021.33.1.40.

Gross Trunnion Failure in the Bipolar Hemiarthroplasty; Raising Concern about Short Trunnion: A Case Report

Affiliations
  • 1Department of Orthopaedic Surgery, Seoul National University Hospital, Seoul, Korea

Abstract

There have been some reports of gross trunnion failure (GTF) in total hip arthroplasty. Here, we report a case of GTF 19 years after bipolar hemiarthroplasty using a 28-mm head with a 14/16 taper bore. Compared to other GTF reports, the current case had some unusual aspects: bipolar hemiarthroplasty, 28-mm head, relatively lateonset, and no apparent findings of metallosis though a severe one was evident. A Computed tomography scout view provided valuable information in evaluating polyethylene, metal head, and neck inside the bipolar cup. The current report suggests a need for concerned regarding short trunnion length which may be associated with GTF.

Keyword

Hip; Arthroplasty; Prosthesis failure