Under or overpressure: an audit of endotracheal cuff pressure monitoring at the tertiary care center

Affiliations
  • 1College of Nursing, JIPMER, Puducherry, India
  • 2Department of Neurosurgery, JIPMER, Puducherry, India

Abstract

Background
Mechanical ventilation is a lifesaving intervention for critically ill patients but can produce the major complication of ventilator-associated pneumonia (VAP). Inappropriately inflated endotracheal tubes cause potential harm due to high or low pressure; this can be prevented through monitoring protocols.
Methods
A cross-sectional study of 348 cuff pressure readings was performed with intubated and mechanically ventilated patients to evaluate the exact proportion of patients in intensive care units (ICUs) where the cuff pressure is optimal and to identify the ICUs where device-based monitoring is available to produce a lower proportion of sub-optimal cuff pressure cases. Every three days, cuff pressure was assessed with a handheld cuff pressure manometer. The corresponding VAP rates of those ICUs were obtained from the hospital infection control department.
Results
Cuff pressure of 40.2% was the lower cutoff for the high category, that of optimal was 35.3%, and the highest cutoff of sub-optimal was 24.4%. This study also showed ICUs that had cuff pressure monitoring devices and protocols. Active measurement protocols had a higher proportion of optimal cuff pressure (58.5%) and a lower proportion of sub-optimal and high cuff pressure (19.5% and 22.0%) compared to ICUs with no device-based monitoring protocols. Furthermore, the VAP rate of ICUs exhibited a weak positive correlation with sub-optimal cuff pressure.
Conclusions
Device-based cuff pressure monitoring is essential in maintaining adequate cuff pressure but often is inadequate, resulting in high readings. Therefore, this study suggests that device-based cuff pressure monitoring be practiced.

Keyword

endo-tracheal cuff pressure; mechanical ventilation; ventilator-associated pneumonia

Reference

1. Magill SS, O'Leary E, Janelle SJ, Thompson DL, Dumyati G, Nadle J, et al. Changes in prevalence of health care-associated infections in U.S. hospitals. N Engl J Med. 2018; 379:1732–44.
Article
2. Haas CF, Eakin RM, Konkle MA, Blank R. Endotracheal tubes: old and new. Respir Care. 2014; 59:933–52.
Article
3. Letvin A, Kremer P, Silver PC, Samih N, Reed-Watts P, Kollef MH. Frequent versus infrequent monitoring of endotracheal tube cuff pressures. Respir Care. 2018; 63:495–501.
Article
4. Akdogan O, Ersoy Y, Kuzucu C, Gedik E, Togal T, Yetkin F. Assessment of the effectiveness of a ventilator associated pneumonia prevention bundle that contains endotracheal tube with subglottic drainage and cuff pressure monitorization. Braz J Infect Dis. 2017; 21:276–81.
Article
5. Hamilton VA, Grap MJ. The role of the endotracheal tube cuff in microaspiration. Heart Lung. 2012; 41:167–72.
Article
6. Rello J, Soñora R, Jubert P, Artigas A, Rué M, Vallés J. Pneumonia in intubated patients: role of respiratory airway care. Am J Respir Crit Care Med. 1996; 154:111–5.
Article
7. Álvarez-Lerma F, Sánchez García M; Task Force of Experts for Project “Zero VAP” in Spain. "The multimodal approach for ventilator-associated pneumonia prevention"-requirements for nationwide implementation. Ann Transl Med. 2018; 6:420.
Article
8. Fu Y, Xi X. Analysis on risk factors of endotracheal cuff under inflation in mechanically ventilated patients. Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2014; 26:870–4.
9. Nseir S, Brisson H, Marquette CH, Chaud P, Di Pompeo C, Diarra M, et al. Variations in endotracheal cuff pressure in intubated critically ill patients: prevalence and risk factors. Eur J Anaesthesiol. 2009; 26:229–34.
Article
10. Nseir S, Zerimech F, Jaillette E, Artru F, Balduyck M. Microaspiration in intubated critically ill patients: diagnosis and prevention. Infect Disord Drug Targets. 2011; 11:413–23.
Article
11. Jaillette E, Zerimech F, De Jonckheere J, Makris D, Balduyck M, Durocher A, et al. Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study. BMC Anesthesiol. 2013; 13:50.
Article
12. Sengupta P, Sessler DI, Maglinger P, Wells S, Vogt A, Durrani J, et al. Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure. BMC Anesthesiol. 2004; 4:8.
Article
13. Ranjan N, Chaudhary U, Chaudhry D, Ranjan KP. Ventilator-associated pneumonia in a tertiary care intensive care unit: analysis of incidence, risk factors and mortality. Indian J Crit Care Med. 2014; 18:200–4.
14. Mukhopadhyay C, Bhargava A, Ayyagari A. Role of mechanical ventilation & development of multidrug resistant organisms in hospital acquired pneumonia. Indian J Med Res. 2003; 118:229–35.
15. Stevens GJ, Warfel JW, Aden JK, Blackwell SD. Intraoperative endotracheal cuff pressure study: how education and availability of manometers help guide safer pressures. Mil Med. 2018; 183:e416–9.
Article
16. Rouzé A, Martin-Loeches I, Nseir S. Airway devices in ventilator-associated pneumonia pathogenesis and prevention. Clin Chest Med. 2018; 39:775–83.
Article
17. Jaillette E, Girault C, Brunin G, Zerimech F, Behal H, Chiche A, et al. Impact of tapered-cuff tracheal tube on microaspiration of gastric contents in intubated critically ill patients: a multicenter cluster-randomized cross-over controlled trial. Intensive Care Med. 2017; 43:1562–71.
Article
18. Alzahrani AR, Al Abbasi S, Abahoussin OK, Al Shehri TO, Al-Dorzi HM, Tamim HM, et al. Prevalence and predictors of out-of-range cuff pressure of endotracheal and tracheostomy tubes: a prospective cohort study in mechanically ventilated patients. BMC Anesthesiol. 2015; 15:147.
Article
19. Huang WM, Huang XA, Du YP, Li LX, Wu FF, Hong SQ, et al. Tapered cuff versus conventional cuff for ventilator-associated pneumonia in ventilated patients: a meta-analysis of randomized controlled trials. Can Respir J. 2019; 2019:7876417.
Article
Full Text Links
  • ACC
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr