Refractive Power Outcomes with an Intraocular Lens with 0.25-diopter Intervals

Affiliations
  • 1Saevit Eye Hospital, Goyang, Korea

Abstract

Purpose
The Eyelike K-flex Aspheric® (Koryoeyetech, Seoul, Korea) is manufactured in 0.25-diopter (D) intervals, which allows the target refractive error after surgery to be achieved. We here evaluate the refractive power outcomes.
Methods
We retrospectively studied 95 eyes of 72 patients who underwent cataract surgery with implantation of the Eyelike K-flex Aspheric®. Refractive error was measured at 1 and 2 months postoperatively (33 eyes of 27 patients) and compared to that of patients fitted with 0.50-D-interval lenses (62 eyes of 49 patients).
Results
At 1 month postoperatively, the mean absolute error between the spherical equivalent and planned value was 0.33 ± 0.28 and 0.41 ± 0.39 D in the 0.25- and 0.50 D-interval lens groups, respectively (p = 0.318). At 2 months postoperatively, the respective values were 0.21 ± 0.15 and 0.34 ± 0.29 D (p = 0.009).
Conclusions
The Eyelike K-flex Aspheric® shows excellent refractive predictability; use of 0.25-D-interval intraocular lenses close to the target refractive power allows the desired spherical equivalent to be achieved.

Keyword

Cataract; Intraocular lens; Refractive error