Polyethylene spinout in the Attune® Cruciate-Retaining Rotating-Platform (CR RP) total knee arthroplasty performed with a cruciate-sacrificing and measuredresection technique

Affiliations
  • 1Department of Trauma and Orthopaedic Surgery, Cork University Hospital/South Infirmary Victoria University Hospital, Wilton, Cork, T12 DC4A, Republic of Ireland
  • 2Department of Orthopaedic Surgery, Bon Secours Hospital Cork, College Road, Cork, Republic of Ireland
  • 3Musgrave Park Hospital, Stockman’s Lane, Belfast, BT9 7JB, Northern Ireland, UK

Abstract

Introduction
Polyethylene (PE) spinout is a known but uncommon complication when using a mobile-bearing (MB) total knee arthroplasty (TKA) design. Sacrificing the posterior cruciate ligament (PCL) is within the manufacturer’s recommendations for the Attune® Cruciate-Retaining Rotating-Platform (CR RP) knee design. Aim: To discuss the potential aetiology and prevention of spinout in the Attune®CR RP knee.
Methods
We used a retrospective radiological review from two centres reporting a higher rate of spinout in the Attune® CR RP knee using a cruciate-sacrificing and measured-resection technique when compared to a gapbalancing technique. Three hundred and thirty-two patients were evaluated over a 3-year period.
Results
There were 8 out of 279 (2.86%) cases of spinout in our first cohort of patients using a measured-resection technique. There were 0 out of 53 cases of spinout in our second cohort of patients where a gap-balancing technique was used. One spinout was reduced closed, the other seven were initially revised to a thicker RP insert of the same design. Of these seven, three underwent a further revision TKA and one patient required a knee fusion/arthrodesis.
Conclusions
This study reports a higher incidence of PE spinout in the Attune® CR RP TKA when a measuredresection technique in combination with PCL resection is performed. We recommend a gap-balancing technique with conservative soft-tissue release if the surgeon is planning to sacrifice the PCL in the Attune® CR RP.

Keyword

Rotating; Platform; Cruciate; Retaining; Gap; Balancing; Attune®; LCS®; Polyethylene; Spinout