J Korean Med Sci.  2016 Nov;31(11):1742-1748. 10.3346/jkms.2016.31.11.1742.

Comparison of Dosimetric Performance among Commercial Quality Assurance Systems for Verifying Pretreatment Plans of Stereotactic Body Radiotherapy Using Flattening-Filter-Free Beams

Affiliations
  • 1Department of Radiation Oncology, Seoul National University Bundang Hospital, Seongnam, Korea.
  • 2Research Institute of Biomedical Engineering, College of Medicine, The Catholic University of Korea, Seoul, Korea. suhsanta@catholic.ac.kr
  • 3Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea, Seoul, Korea.
  • 4Department of Radiation Oncology, SAM Anyang Hospital, Anyang, Korea.

Abstract

The purpose of this study was to compare the performance of different commercial quality assurance (QA) systems for the pretreatment verification plan of stereotactic body radiotherapy (SBRT) with volumetric arc therapy (VMAT) technique using a flattening-filter-free beam. The verification for 20 pretreatment cancer patients (seven lung, six spine, and seven prostate cancers) were tested using three QA systems (EBT3 film, I'mRT MatriXX array, and MapCHECK). All the SBRT-VMAT plans were optimized in the Eclipse (version 11.0.34) treatment planning system (TPS) using the Acuros XB dose calculation algorithm and were delivered to the Varian TrueBeam® accelerator equipped with a high-definition multileaf collimator. Gamma agreement evaluation was analyzed with the criteria of 2% dose difference and 2 mm distance to agreement (2%/2 mm) or 3%/3 mm. The highest passing rate (99.1% for 3%/3 mm) was observed on the MapCHECK system while the lowest passing rate was obtained on the film. The pretreatment verification results depend on the QA systems, treatment sites, and delivery beam energies. However, the delivery QA results for all QA systems based on the TPS calculation showed a good agreement of more than 90% for both the criteria. It is concluded that the three 2D QA systems have sufficient potential for pretreatment verification of the SBRT-VMAT plan.

Keyword

Pretreatment Verification; Delivery Quality Assurance; Stereotactic Body Radiotherapy

MeSH Terms

Algorithms
Bone Neoplasms/radiotherapy
Humans
Lung Neoplasms/radiotherapy
Male
Prostatic Neoplasms/radiotherapy
*Quality Assurance, Health Care
Radiometry/instrumentation/*methods
*Radiosurgery
Radiotherapy Dosage
Retrospective Studies

Figure

  • Fig. 1 Three commercial QA systems set up for measuring pretreatment plan of SBRT-VMAT. (A) EBT3 film using cylindrical acryl phantom. (B) I’mRT MatriXX using the MultiCube phantom. (C) MapCHECK using the MapPHAN phantom.

  • Fig. 2 Example of gamma evaluation results between the three QA systems and TPS. The measured dose distribution, the calculated dose distribution and the gamma evaluation image for (A) EBT3 film, (B) I’mRT MatriXX, and (C) MapCHECK.

  • Fig. 3 Gamma passing rate of MapCHECK and MatriXX for static field of 6-MV FFF beam with various dose rates and different field sizes.


Reference

1. Vieillevigne L, Molinier J, Brun T, Ferrand R. Gamma index comparison of three VMAT QA systems and evaluation of their sensitivity to delivery errors. Phys Med. 2015; 31:720–725.
2. Thomas SD, Mackenzie M, Field GC, Syme AM, Fallone BG. Patient specific treatment verifications for helical tomotherapy treatment plans. Med Phys. 2005; 32:3793–3800.
3. Ezzell GA, Galvin JM, Low D, Palta JR, Rosen I, Sharpe MB, Xia P, Xiao Y, Xing L, Yu CX. Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. Med Phys. 2003; 30:2089–2115.
4. Kim YL, Chung JB, Kim JS, Lee JW, Choi KS. Comparison of the performance between portal dosimetry and a commercial two-dimensional array system on pretreatment quality assurance for volumetric-modulated arc and intensity-modulated radiation therapy. J Korean Phys Soc. 2014; 64:1207–1212.
5. Kim TH, Oh S, Kim MJ, Jung WG, Chung JB, Kim JS, Kim S, Suh TS. Efficiency study of 2D diode array detector for IMRT quality assurance. Korean J Med Phys. 2011; 22:61–66.
6. Lee JW, Hong S, Kim YL, Choi KS, Chung JB, Lee DH, Suh TS. Dosimetric characterization of an ion chamber matrix for intensity modulated radiation therapy quality assurance. Korean J Med Phys. 2006; 17:131–135.
7. Son J, Baek T, Lee B, Shin D, Park SY, Park J, Lim YK, Lee SB, Kim J, Yoon M. A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy. Radiol Oncol. 2015; 49:307–313.
8. Elith CA, Dempsey SE, Cao F, Farshadi A, Warren-Forward HM. The quality assurance of volumetric modulated arc therapy (VMAT) plans for early stage prostate cancer: a technical note. J Med Radiat Sci. 2014; 61:261–266.
9. Kry SF, Vassiliev ON, Mohan R. Out-of-field photon dose following removal of the flattening filter from a medical accelerator. Phys Med Biol. 2010; 55:2155–2166.
10. Scorsetti M, Alongi F, Castiglioni S, Clivio A, Fogliata A, Lobefalo F, Mancosu P, Navarria P, Palumbo V, Pellegrini C, et al. Feasibility and early clinical assessment of flattening filter free (FFF) based stereotactic body radiotherapy (SBRT) treatments. Radiat Oncol. 2011; 6:113.
11. Cashmore J, Ramtohul M, Ford D. Lowering whole-body radiation doses in pediatric intensity-modulated radiotherapy through the use of unflattened photon beams. Int J Radiat Oncol Biol Phys. 2011; 80:1220–1227.
12. Kragl G, Baier F, Lutz S, Albrich D, Dalaryd M, Kroupa B, Wiezorek T, Knöös T, Georg D. Flattening filter free beams in SBRT and IMRT: dosimetric assessment of peripheral doses. Z Med Phys. 2011; 21:91–101.
13. Chung JB, Kim JS, Eom KY, Kim IA, Kang SW, Lee JW, Kim JY, Suh TS. Comparison of VMAT-SABR treatment plans with flattening filter (FF) and flattening filter-free (FFF) beam for localized prostate cancer. J Appl Clin Med Phys. 2015; 16:5728.
14. Kim YL, Chung JB, Kim JS, Lee JW, Kim JY, Kang SW, Suh TS. Dosimetric comparison of a 6-MV flattening-filter and a flattening-filter-free beam for lung stereotactic ablative radiotherapy treatment. J Korean Phys Soc. 2015; 67:1672–1678.
15. Chung JB, Eom KY, Kim IA, Kim JS, Lee JW, Hong S, Kim YL, Park BM, Kang SW, Suh TS. Comparison of anisotropic analytic algorithm plan and acuros XB plan for lung stereotactic ablative radiotherapy using flattening filter-free beams. Prog Med Phys. 2014; 25:210–217.
16. Lang S, Reggiori G, Puxeu Vaquee J, Calle C, Hrbacek J, Klock S, Scorsetti M, Cozzi L, Mancosu P. Pretreatment quality assurance of flattening filter free beams on 224 patients for intensity modulated plans: a multicentric study. Med Phys. 2012; 39:1351–1356.
17. IBA Dosimetry (DE). MatriXX evolution system: the solution for rotational treatment QA. accessed on 15 June 2015. Available at http://www.iba-dosimetry.com.
18. IBA Dosimetry (DE). I'mRT MatriXX: the new standard in 2D IMRT pre-treatment verification. accessed on 15 June 2015. Available at http://www.iba-dosimetry.
19. Gloi AM, Buchana RE, Zuge CL, Goettler AM. RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys. 2011; 12:3251.
20. Jursinic PA, Sharma R, Reuter J. MapCHECK used for rotational IMRT measurements: step-and-shoot, TomoTherapy, RapidArc. Med Phys. 2010; 37:2837–2846.
21. Alvarez-Moret J, Pohl F, Koelbl O, Dobler B. Evaluation of volumetric modulated arc therapy (VMAT) with Oncentra MasterPlan® for the treatment of head and neck cancer. Radiat Oncol. 2010; 5:110.
22. Nelms BE, Simon JA. A survey on planar IMRT QA analysis. J Appl Clin Med Phys. 2007; 8:2448.
23. Iftimia I, Cirino ET, Xiong L, Mower HW. Quality assurance methodology for Varian RapidArc treatment plans. J Appl Clin Med Phys. 2010; 11:3164.
24. Bailey DW, Kumaraswamy L, Bakhtiari M, Malhotra HK, Podgorsak MB. EPID dosimetry for pretreatment quality assurance with two commercial systems. J Appl Clin Med Phys. 2012; 13:3736.
25. Shim SJ, Shim JB, Lee SH, Min CK, Cho KH, Shin DO, Choi JH, Park SI, Cho SJ. Quality assurance of volumetric modulated arc therapy for elekta synergy. Korean J Med Phys. 2012; 23:33–41.
26. Rinaldin G, Perna L, Agnello G, Pallazzi G, Cattaneo GM, Fiorino C, Calandrino R. Quality assurance of rapid arc treatments: performances and pre-clinical verifications of a planar detector (MapCHECK2). Phys Med. 2014; 30:184–190.
27. Masi L, Casamassima F, Doro R, Francescon P. Quality assurance of volumetric modulated arc therapy: evaluation and comparison of different dosimetric systems. Med Phys. 2011; 38:612–621.
28. Hussein M, Rowshanfarzad P, Ebert MA, Nisbet A, Clark CH. A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems. Radiother Oncol. 2013; 109:370–376.
29. Hussein M, Adams EJ, Jordan TJ, Clark CH, Nisbet A. A critical evaluation of the PTW 2D-ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification. J Appl Clin Med Phys. 2013; 14:4460.
30. Herzen J, Todorovic M, Cremers F, Platz V, Albers D, Bartels A, Schmidt R. Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine. Phys Med Biol. 2007; 52:1197–1208.