Anesth Pain Med.  2019 Oct;14(4):423-428. 10.17085/apm.2019.14.4.423.

Intraoperative continuous noninvasive hemoglobin monitoring in patients with placenta previa undergoing cesarean section: a prospective observational study

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, Seoul National University Bundang Hospital, Seongnam, Korea. hsknana@gmail.com

Abstract

BACKGROUND
Obstetric patients with placenta previa are at risk for sever peripartum hemorrhage. Early detection of anemia and proper transfusion strategy are important for the management of obstetric hemorrhage. In this study, we assessed the utility and accuracy of noninvasive hemoglobin (SpHb) monitoring in patients with placenta previa during cesarean section.
METHODS
Parturients diagnosed with placenta previa and scheduled for cesarean section under spinal anesthesia were enrolled. SpHb and laboratory Hb (Lab-Hb) were measured during surgery as primary outcomes.
RESULTS
Seventy-four pairs of SpHb and Lab-Hb were collected from 39 patients. The correlation coefficient was 0.877 between SpHb and Lab-Hb (P < 0.001). The Bland-Altman plot showed a mean difference ± SD of 0.3 ± 0.8 g/dl between noninvasive Hb and Lab-Hb, and the limits of agreement were −1.2 to 1.8 g/dl. The magnitude of the difference between SpHb and Lab-Hb was < 0.5 g/dl in 64.9%; however, it was > 1.5 g/dl in 10.8%.
CONCLUSIONS
SpHb monitoring had a good correlation with Lab-Hb. A small mean difference between SpHb and lab-Hb might not be clinically significant; however, the limits of agreements were not narrow. In particular, SpHb could be overestimated in the anemic population. Based on our results, further studies investigating the accuracy and precision of SpHb monitoring should be performed in parturients presenting Hb below 10 g/dl.

Keyword

Cesarean section; Laboratory hemoglobin; Noninvasive hemoglobin

MeSH Terms

Anemia
Anesthesia, Spinal
Cesarean Section*
Female
Hemorrhage
Humans
Observational Study*
Peripartum Period
Placenta Previa*
Placenta*
Pregnancy
Prospective Studies*

Figure

  • Fig. 1 Change of noninvasive hemoglobin (SpHb) and perfusion index (Pi). OP: operation.

  • Fig. 2 Correlation between the laboratory hemoglobin (Lab-Hb) and noninvasive hemoglobin (SpHb). The correlation coefficient was 0.877 between the SpHb and Lab-Hb.

  • Fig. 3 Bland–Altman analysis for laboratory hemoglobin (Lab-Hb) and noninvasive hemoglobin (SpHb). The mean difference (one horizontal solid line) between the SpHb and Lab-Hb was 0.3 g/dl and the limits of agreement (two horizontal dash lines) were −1.2 to 1.8 g/dl.


Reference

1. García-Soler P, Camacho Alonso JM, González-Gómez JM, Mi-lano-Manso G. Noninvasive hemoglobin monitoring in critically ill pediatric patients at risk of bleeding. Med Intensiva. 2017; 41:209–15. DOI: 10.1016/j.medine.2016.06.005. PMID: 28034464.
2. Xu T, Yang T, Kim JB, Romig MC, Sapirstein A, Winters BD. Evaluation of noninvasive hemoglobin monitoring in surgical critical care patients. Crit Care Med. 2016; 44:344–52. DOI: 10.1097/CCM.0000000000001634. PMID: 26937862.
3. Martin JR, Camp CL, Stitz A, Young EY, Abdel MP, Taunton MJ, et al. Noninvasive hemoglobin monitoring:a rapid, reliable, and cost-effective method following total joint replacement. J Bone Joint Surg Am. 2016; 98:349–55. DOI: 10.2106/JBJS.O.00820. PMID: 26935456.
4. Galvagno SM Jr, Hu P, Yang S, Gao C, Hanna D, Shackelford S, et al. Accuracy of continuous noninvasive hemoglobin monitoring for the prediction of blood transfusions in trauma patients. J Clin Monit Comput. 2015; 29:815–21. DOI: 10.1007/s10877-015-9671-1. PMID: 25753142.
5. Phillips MR, Khoury AL, Bortsov AV, Marzinsky A, Short KA, Cairns BA, et al. A noninvasive hemoglobin monitor in the pediatric intensive care unit. J Surg Res. 2015; 195:257–62. DOI: 10.1016/j.jss.2014.12.051. PMID: 25724765. PMCID: PMC5892184.
6. Moore LJ, Wade CE, Vincent L, Podbielski J, Camp E, Junco DD, et al. Evaluation of noninvasive hemoglobin measurements in trauma patients. Am J Surg. 2013; 206:1041–7. DOI: 10.1016/j.amjsurg.2013.08.012. PMID: 24157223.
7. Bhat A, Upadhyay A, Jaiswal V, Chawla D, Singh D, Kumar M, et al. Validity of non-invasive point-of-care hemoglobin estimation in healthy and sick children-a method comparison study. Eur J Pediatr. 2016; 175:171–9. DOI: 10.1007/s00431-015-2602-9. PMID: 26286814.
8. Bourke M, Cooley SM, McMorrow R, McAuliffe FM. Noninvasive hemoglobin measurement in pregnancy. Int J Gynaecol Obstet. 2014; 125:78–9. DOI: 10.1016/j.ijgo.2013.10.014. PMID: 24524913.
9. Pyregov AV, Ovechkin AIu, Petrov SV. Noninvasive total hemoglobin monitoring based on multiwave spectrophotometry in obstetrics and gynecology. Anesteziol Reanimatol. 2012; 6:36–9. PMID: 23662518.
10. Huang PH, Shih BF, Tsai YF, Chung PC, Liu FC, Yu HP, et al. Accuracy and trending of continuous noninvasive hemoglobin monitoring in patients undergoing liver transplantation. Transplant Proc. 2016; 48:1067–70. DOI: 10.1016/j.transproceed.2015.12.121. PMID: 27320558.
11. Suehiro K, Joosten A, Alexander B, Cannesson M. Continuous noninvasive hemoglobin monitoring:ready for prime time? Curr Opin Crit Care. 2015; 21:265–70. DOI: 10.1097/MCC.0000000000000197. PMID: 25827587.
12. Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011; 113:1396–402. DOI: 10.1213/ANE.0b013e318230b425. PMID: 21965372.
13. Park YH, Lee JH, Song HG, Byon HJ, Kim HS, Kim JT. The accuracy of noninvasive hemoglobin monitoring using the radical-7 pulse CO-Oximeter in children undergoing neurosurgery. Anesth Analg. 2012; 115:1302–7. DOI: 10.1213/ANE.0b013e31826b7e38. PMID: 23144437.
14. Kim SH, Choi JM, Kim HJ, Choi SS, Choi IC. Continuous noninvasive hemoglobin measurement is useful in patients undergoing double-jaw surgery. J Oral Maxillofac Surg. 2014; 72:1813–9. DOI: 10.1016/j.joms.2014.03.011. PMID: 24813777.
15. Butwick A, Hilton G, Carvalho B. Non-invasive haemoglobin measurement in patients undergoing elective Caesarean section. Br J Anaesth. 2012; 108:271–7. DOI: 10.1093/bja/aer373. PMID: 22116296.
16. Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices:a comparison of Radical-7TM pulse co-oximetry, HemoCue™and laboratory haemoglobin measurements in obstetric patients. Anaesthesia. 2013; 68:40–5. DOI: 10.1111/anae.12039. PMID: 23088815.
17. Awada WN, Mohmoued MF, Radwan TM, Hussien GZ, Elkady HW. Continuous and noninvasive hemoglobin monitoring reduces red blood cell transfusion during neurosurgery:a prospective cohort study. J Clin Monit Comput. 2015; 29:733–40. DOI: 10.1007/s10877-015-9660-4. PMID: 25649717. PMCID: PMC4621711.
18. Gamal M, Abdelhamid B, Zakaria D, Dayem OAE, Rady A, Fawzy M, et al. Evaluation of noninvasive hemoglobin monitoring in trauma patients with low hemoglobin levels. Shock. 2018; 49:150–3. DOI: 10.1097/SHK.0000000000000949. PMID: 28727608.
19. Saito J, Kitayama M, Oishi M, Kudo T, Sawada M, Hashimoto H, et al. The accuracy of non-invasively continuous total hemoglobin measurement by pulse CO-Oximetry undergoing acute normovolemic hemodilution and reinfusion of autologous blood. J Anesth. 2015; 29:29–34. DOI: 10.1007/s00540-014-1863-1. PMID: 24972855.
20. Baulig W, Seifert B, Spahn DR, Theusinger OM. Accuracy of noninvasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients. J Clin Monit Comput. 2017; 31:177–85. DOI: 10.1007/s10877-015-9816-2. PMID: 26686689.
21. Park SG, Lee OH, Park YH, Shin HY, Kang H, Baek CW, et al. The changes of non-invasive hemoglobin and perfusion index of Pulse CO-Oximetry during induction of general anesthesia. Korean J Anesthesiol. 2015; 68:352–7. DOI: 10.4097/kjae.2015.68.4.352. PMID: 26257847. PMCID: PMC4524933.
22. Miller RD, Ward TA, McCulloch CE, Cohen NH. Does a digital regional nerve block improve the accuracy of noninvasive hemoglobin monitoring? J Anesth. 2012; 26:845–50. DOI: 10.1007/s00540-012-1452-0. PMID: 22850903.
23. Nguyen BV, Vincent JL, Nowak E, Coat M, Paleiron N, Gouny P, et al. The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg. 2011; 113:1052–7. DOI: 10.1213/ANE.0b013e31822c9679. PMID: 21918163.
24. Yang ZW, Yang SH, Chen L, Qu J, Zhu J, Tang Z. Comparison of blood counts in venous, fingertip and arterial blood and their measurement variation. Clin Lab Haematol. 2001; 23:155–9. DOI: 10.1046/j.1365-2257.2001.00388.x. PMID: 11553055.
25. Carabini LM, Navarre WJ, Ault ML, Bebawy JF, Gupta DK. A comparison of hemoglobin measured by co-oximetry and central laboratory during major spine fusion surgery. Anesth Analg. 2015; 120:60–5. DOI: 10.1213/ANE.0000000000000418. PMID: 25185592.
Full Text Links
  • APM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr