Ultrasonography.  2019 Oct;38(4):345-354. 10.14366/usg.19001.

Inter-platform reproducibility of liver stiffness measured with two different point shear wave elastography techniques and 2-dimensional shear wave elastography using the comb-push technique

Affiliations
  • 1Department of Radiology, Pusan National University Yangsan Hospital, Yangsan, Korea.
  • 2Department of Radiology, Gachon University Gil Medical Center, Incheon, Korea.
  • 3Department of Radiology, Seoul National University Hospital, Seoul, Korea. jmsh@snu.ac.kr
  • 4Department of Radiology, Seoul National University College of Medicine, Seoul, Korea.
  • 5Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Korea.

Abstract

PURPOSE
The purpose of this study was to compare the technical success rate and reliability of measurements made using three shear wave elastography (SWE) techniques and to assess the inter-platform reproducibility of the resultant liver stiffness measurements.
METHODS
This prospective study included 54 patients with liver disease. Liver stiffness (LS) measurements were obtained using 2-point SWE techniques (Virtual Touch Quantification and S-Shearwave) and 2-dimensional (2D) SWE, with transient elastography (TE) serving as the reference standard. The technical success rates and measurement reliability of the three techniques were compared. LS values measured using the three SWE techniques and TE were compared using Spearman correlation coefficients and 95% Bland-Altman limits of agreement. Intra-class correlation coefficients (ICC) were used to analyze the inter-platform reproducibility of LS measurements.
RESULTS
The three SWE techniques and TE showed similar technical success rates (P=0.682) but demonstrated significant differences in the reliability of LS measurements (P=0.006) and mean LS measurements (P<0.001). Despite strong correlations (r=0.73-0.94) between SWE systems, various degrees of inter-platform reproducibility (ICC, 0.58-0.92) were observed for the three SWE techniques. The best agreement was observed between S-Shearwave and TE (ICC, 0.92), and the worst agreement was observed between 2D-SWE and TE (ICC, 0.58). In the Bland-Altman analysis, a tendency toward lower LS values with the three SWE techniques than with TE in patients with F3 and F4 disease was observed.
CONCLUSION
Significant inter-system variability was observed in LS measurements made using the three SWE techniques. Therefore, LS values measured using different SWE techniques should not be used interchangeably for longitudinal follow-up.

Keyword

Liver fibrosis; Liver stiffness; Ultrasound elastography; Shear wave elastography

MeSH Terms

Elasticity Imaging Techniques*
Follow-Up Studies
Humans
Liver Cirrhosis
Liver Diseases
Liver*
Prospective Studies