Non-traumatic Ceramic Head Fracture in Total Hip Arthroplasty with Ceramic-on-Ceramic Articulation at Postoperative 16th Years

Affiliations
  • 1Department of Orthopaedics and Traumatology, Yeditepe University Faculty of Medicine, Istanbul, Turkey. onurkocadal@gmail.com

Abstract

Ceramic articulation has gained popularity in total hip arthroplasty (THA); however, one of the most important concerns about the use of ceramic materials is the potential for breakage. Importantly, almost all ceramic fractures occur within the first six years after surgery. Here, we present a case report of a 52-year-old female patient who experienced an atraumatic ceramic head fracture at the 16th year after surgery. Key parameters that may be associated with ceramic head fracture are considered to be the use of a 28-mm head and high body mass index (BMI); the patient described here had a BMI of 34.3 kg/m2. Surgical treatment of the case included removal of the broken ceramic head and replacement with an alumina matrix composite ceramic head. At the 12th month follow-up visit, the patient was mobilized independently and could perform her daily activities. Ceramic head fracture after THA may occur even in late periods after surgery. For the treatment of fractures of ceramic material, the fractured component must be removed together with extensive soft tissue excision to ensure that no residual fragments are left behind.

Keyword

Ceramics; Arthroplasty; Hip; Hip prosthesis

MeSH Terms

Aluminum Oxide
Arthroplasty
Arthroplasty, Replacement, Hip*
Body Mass Index
Ceramics*
Female
Follow-Up Studies
Head*
Hip
Hip Prosthesis
Humans
Middle Aged
Aluminum Oxide

Figure

  • Fig. 1 Pelvic anteroposterior radiography reveals the non-centralized position of the ceramic head in the acetabulum on the right side.

  • Fig. 2 Computed tomographic evaluation reveals fragmentation in the ceramic head (A). Acetabular cup anteversion was measured as 14° (B).

  • Fig. 3 Intraoperative image of the broken ceramic head.


Reference

1. Pulliam IT, Trousdale RT. Fracture of a ceramic femoral head after a revision operation. A case report. J Bone Joint Surg Am. 1997; 79:118–121.
Article
2. Hwang DS, Kim YM, Lee CH. Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty. 2007; 22:468–471.
Article
3. Kumar N, Sharma V. Error in surgical technique causing ceramic acetabular liner fracture in primary total hip arthroplasty - a report of two cases. J Arthroscop Joint Surg. 2016; 3:75–77.
Article
4. Tucker D, Acharya M. Unusual prosthetic femoral head fracture in total hip arthroplasty: ceramic on polyethylene articulation. BMJ Case Rep. 2014; 2014:bcr2013202685.
Article
5. Beckmann NA, Gotterbarm T, Innmann MM, et al. Long-term durability of alumina ceramic heads in THA. BMC Musculoskelet Disord. 2015; 16:249.
Article
6. Lee GC, Kim RH. Incidence of modern alumina ceramic and alumina matrix composite femoral head failures in nearly 6 million hip implants. J Arthroplasty. 2017; 32:546–551.
Article
7. Knahr K, Böhler M, Frank P, Plenk H, Salzer M. Survival analysis of an uncemented ceramic acetabular component in total hip replacement. Arch Orthop Trauma Surg. 1987; 106:297–300.
Article
8. Ha YC, Kim SY, Kim HJ, Yoo JJ, Koo KH. Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty. Clin Orthop Relat Res. 2007; 458:106–110.
Article
9. Morlock MM, Witt F, Bishop N, et al. Wear of a composite ceramic head caused by liner fracture. Orthopedics. 2014; 37:e653–e656.
Article
10. Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002; (401):149–161.
Article
11. Boutin P. Total arthroplasty of the hip by fritted aluminum prosthesis. Experimental study and 1st clinical applications. Rev Chir Orthop Reparatrice Appar Mot. 1972; 58:229–246.
12. Nho JH, Park JS, Song US, Kim WJ, Suh YS. Ceramic head fracture in ceramic-on-polyethylene total hip arthroplasty. Yonsei Med J. 2013; 54:1550–1553.
Article
13. Min BW, Song KS, Kang CH, Bae KC, Won YY, Lee KY. Delayed fracture of a ceramic insert with modern ceramic total hip replacement. J Arthroplasty. 2007; 22:136–139.
Article
14. Traina F, De Fine M, Di Martino A, Faldini C. Fracture of ceramic bearing surfaces following total hip replacement: a systematic review. Biomed Res Int. 2013; 2013:157247.
Article
15. Oak SR, Frangiamore SJ, Higuera Rueda CA, Klika A, Molloy R. Atraumatic delta head fracture in a ceramic-on-polyethylene total hip arthroplasty. J Hip Surg. J Hip Surg. 2017; 1:118–120.
16. Arenas A, Tejero A, Garbayo A, Arias C, Barbadillo J, Lecumberri J. Ceramic femoral head fractures in total hip replacement. Int Orthop. 1999; 23:351–352.
Article
17. Howard DP, Wall PDH, Fernandez MA, Parsons H, Howard PW. Ceramic-on-ceramic bearing fractures in total hip arthroplasty: an analysis of data from the National Joint Registry. Bone Joint J. 2017; 99-B:1012–1019.
18. Migaud H, Putman S, Kern G, et al. Do the reasons for ceramic-on-ceramic revisions differ from other bearings in total hip arthroplasty? Clin Orthop Relat Res. 2016; 474:2190–2199.
Article
19. Hannouche D, Zingg M, Miozzari H, Nizard R, Lübbeke A. Third-generation pure alumina and alumina matrix composites in total hip arthroplasty: what is the evidence? EFORT Open Rev. 2018; 3:7–14.
Article
Full Text Links
  • HP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr