Korean J Gastroenterol.  2018 Jun;71(6):319-323. 10.4166/kjg.2018.71.6.319.

Efficacy of Peracetic Acid (EndoPA®) for Disinfection of Endoscopes

Affiliations
  • 1Department of Internal Medicine, St. Vincent's Hospital, The Catholic University of Korea, Suwon, Korea. drmaloman@catholic.ac.kr

Abstract

BACKGROUND/AIMS
We aimed to investigate the efficacy of peracetic acid (EndoPA®; Firson Co., Ltd., Cheonan, Korea) in disinfecting endoscopes.
METHODS
We prospectively investigated the gastroscopes (Part I) utilized in 100 gastroscopic examinations and colonoscopes (Part II) utilized in 30 colonoscopic examinations after disinfecting them with 0.2% peracetic acid (EndoPA®; Firson Co., Ltd.). These instruments had been collected consecutively throughout the study period. We reprocessed and disinfected the endoscopes according to the guidelines for cleaning and disinfecting gastrointestinal endoscopes laid down by the Korean Society of Gastrointestinal Endoscopy in 2017. Three culture samples were obtained from each examination, based on different sampling methods. The primary outcome was a positive culture rate.
RESULTS
In Part I of our study, two of 300 samples were positive. The culture positive rate after disinfection was 0.7% (2/300). The culture positive rate was not significantly different based on the exposure time to EndoPA® or the age of the scopes (p=0.7 or 0.2, respectively). In Part II of our study, all samples (n=90) were negative.
CONCLUSIONS
We conclude that 0.2% peracetic acid (EndoPA®) appears to be a good disinfectant for both gastroscopes and colonoscopes.

Keyword

Peracetic acid; Endoscope; Disinfection; Efficacy

MeSH Terms

Chungcheongnam-do
Colonoscopes
Disinfection*
Endoscopes*
Endoscopes, Gastrointestinal
Endoscopy, Gastrointestinal
Gastroscopes
Peracetic Acid*
Prospective Studies
Peracetic Acid

Reference

References

1. Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993; 118:117–128.
Article
2. Langenberg W, Rauws EA, Oudbier JH, Tytgat GN. Patient-to-patient transmission of Campylobacter pylori infection by fiberoptic gastroduodenoscopy and biopsy. J Infect Dis. 1990; 161:507–511.
Article
3. Akamatsu T, Tabata K, Hironga M, Kawakami H, Uyeda M. Transmission of Helicobacter pylori infection via flexible fiberoptic endoscopy. Am J Infect Control. 1996; 24:396–401.
Article
4. Centers for Disease Control (CDC). Nosocomial infection and pseudoinfection from contaminated endoscopes and broncho-scopes–Wisconsin and Missouri. MMWR Morb Mortal Wkly Rep. 1991; 40:675–678.
5. ASGE Quality Assurance In Endoscopy Committee. Petersen BT, Chennat J, et al. Multisociety guideline on reprocessing flexible gastrointestinal endoscopes: 2011. Gastrointest Endosc. 2011; 73:1075–1084.
Article
6. Beilenhoff U, Neumann CS, Rey JF, et al. ESGE-ESGENA guideline: cleaning and disinfection in gastrointestinal endoscopy. Endoscopy. 2008; 40:939–957.
Article
7. Kim BW, Son BK, Kim WH, et al. Korean Society of Gastrointestinal Endoscopy (KSGE) guidelines for endoscope reprocessing. Korean J Med. 2017; 92:239–244.
Article
8. Cheung RJ, Ortiz D, DiMarino AJ Jr. GI endoscopic reprocessing practices in the United States. Gastrointest Endosc. 1999; 50:362–368.
Article
9. Ruddy M, Kibbler CC. Endoscopic decontamination: an audit and practical review. J Hosp Infect. 2002; 50:261–268.
Article
10. Park S, Jang JY, Koo JS, et al. A review of current disinfectants for gastrointestinal endoscopic reprocessing. Clin Endosc. 2013; 46:337–341.
Article
11. Kim JB, Han DS, Lee HL, et al. The value of peracetic acid (SCOTELIN®) for endoscope disinfection. Korean J Gastrointest Endosc. 2004; 28:284–290.
12. Chenjiao W, Hongyan Z, Qing G, Xiaoqi Z, Liying G, Ying F. In-use evaluation of peracetic acid for high-level disinfection of endoscopes. Gastroenterol Nurs. 2016; 39:116–120.
Article
13. Seo HI, Lee DS, Yoon EM, et al. Comparison of the efficacy of disinfectants in automated endoscope reprocessors for colonoscopes: tertiary amine compound (Sencron2®) versus or-tho-phthalaldehyde (Cidex®OPA). Intest Res. 2016; 14:178–182.
Article
14. Bradley CR, Babb JR, Ayliffe GA. Evaluation of the steris system 1 peracetic acid endoscope processor. J Hosp Infect. 1995; 29:143–151.
Article
15. Crow S. Peracetic acid sterilization: a timely development for a busy healthcare industry. Infect Control Hosp Epidemiol. 1992; 13:111–113.
Article
16. Sagripanti JL, Bonifacino A. Comparative sporicidal effect of liquid chemical germicides on three medical devices contaminated with spores of Bacillus subtilis. Am J Infect Control. 1996; 24:364–371.
Article
17. Cooke RP, Goddard SV, Whymant-Morris A, Sherwood J, Chatterly R. An evaluation of Cidex OPA (0.55% ortho-phthalaldehyde) as an alternative to 2% glutaraldehyde for high-level disinfection of endoscopes. J Hosp Infect. 2003; 54:226–231.
Article
18. Sokol WN. Nine episodes of anaphylaxis following cystoscopy caused by Cidex OPA (ortho-phthalaldehyde) high-level disinfectant in 4 patients after cytoscopy. J Allergy Clin Immunol. 2004; 114:392–397.
Article
19. Rutala WA, Weber DJ. Surface disinfection: should we do it? J Hosp Infect. 2001; 48(Suppl A):S64–68.
Article
20. Zhang X, Kong J, Tang P, et al. Current status of cleaning and disinfection for gastrointestinal endoscopy in China: a survey of 122 endoscopy units. Dig Liver Dis. 2011; 43:305–308.
Article
Full Text Links
  • KJG
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr