Lab Med Online.  2017 Oct;7(4):170-175. 10.3343/lmo.2017.7.4.170.

Evaluation of the Automated Blood Bank Systems IH-500 and VISION Max for ABO-RhD Blood Typing and Unexpected Antibody Screening

Affiliations
  • 1Department of Laboratory Medicine, Chungnam National University College of Medicine, Daejeon, Korea. kckwon00@naver.com
  • 2Department of Laboratory Medicine, Ajou University School of Medicine, Suwon, Korea.

Abstract

BACKGROUND
The use of automated systems for pre-transfusion tests is increasing in an attempt to reduce workload and the impact of human errors in blood banks. We evaluated the clinical performance of the automated blood bank systems IH-500 (Bio-Rad Laboratories, Switzerland) and VISION Max (Ortho-Clinical Diagnostics, USA) for ABO-RhD blood typing and unexpected antibody screening.
METHODS
ABO-RhD blood typing was performed for 410 samples, and antibody screening was performed for 332 samples, including 15 antibody-positive samples. The results obtained from the two automated instruments were compared with those obtained using manual methods for ABO-RhD blood typing and a semiautomated method (DiaMed-ID system) for antibody screening. Additionally, both instruments were evaluated in terms of concordance rates, sensitivity, and carryover.
RESULTS
The concordance rate of the ABO-RhD blood typing results between the manual methods and the two automated instruments was 100%. For antibody screening tests, the concordance rates between the semiautomated method (DiaMed-ID system) and the automated methods were 100% and 99.7% for the IH-500 and VISION Max instruments, respectively. The sole discrepant result was obtained for a sample identified as antibody-positive only on the VISION Max; the antibody was identified as anti-Le(a). The overall sensitivity of the two automated instruments was the same as or higher than that of the semiautomated method. Carryover was not observed in antibody screening.
CONCLUSIONS
The IH-500 and VISION Max instruments showed reliable results for ABO-RhD blood typing and unexpected antibody screening, and can be used clinically, with confidence, for pre-transfusion tests in the blood bank.

Keyword

Blood bank; Automation; IH-500; VISION Max

MeSH Terms

Automation
Blood Banks*
Blood Grouping and Crossmatching*
Humans
Mass Screening*
Methods

Cited by  1 articles

Evaluation of the Automated Cross-Matching Instrument, ORTHO VISION, for Use in Blood Banks
Mosae Koo, Jinsook Lim, Seon Yung Kim, Ji Myung Kim, Sun Hoe Koo, Gye Cheol Kwon
Lab Med Online. 2019;9(4):218-223.    doi: 10.3343/lmo.2019.9.4.218.


Reference

1. Han KS, Park KU, Song EY. Transfusion medicine. 4th ed. Seoul: Korea Medical Book Publisher;2014. p. 269–277.
2. Malomgré W, Neumeister B. Recent and future trends in blood group typing. Anal Bioanal Chem. 2009; 393:1443–1451.
Article
3. Shin KH, Kim HH, Chang CL, Lee EY. Economic and workflow analysis of a blood bank automated system. Ann Lab Med. 2013; 33:268–273.
Article
4. Shin SY, Kwon KC, Koo SH, Park JW, Ko CS, Song JH, et al. Evaluation of two automated instruments for pre-transfusion testing: AutoVue Innova and Techno TwinStation. Korean J Lab Med. 2008; 28:214–220.
Article
5. Song YK, An TK, Hwang JY, Shim HE, Lee DH, Kong SY. Comparison of effectiveness between blood bank automation system and manual method for ABO-RhD blood typing and antibody screening test in a single center. Korean J Blood Transfus. 2014; 25:93–98.
6. Garratty G. Advances in red blood cell immunology 1960 to 2009. Transfusion. 2010; 50:526–535.
7. Bajpai M, Kaur R, Gupta E. Automation in immunohematology. Asian J Transfus Sci. 2012; 6:140.
Article
8. Clinical and Laboratory Standards Institute (CLSI). CLSI document EP12-A2. Wayne, PA: Clinical and Laboratory Standards Institute;2008.
9. Park Y, Lim J, Ko YH, Kwon K, Koo S, Kim J. Evaluation of IH-1000 for automated ABO-Rh typing and irregular antibody screening. Korean J Blood Transfus. 2012; 23:127–135.
10. Park JS, Park CM, Chang HE, Kim MJ, Park KU, Song J, et al. Evaluation of the performance of the DG Gel test for unexpected antibody screening and identification. Korean J Blood Transfus. 2008; 19:49–56.
11. Cid J, Nogues N, Montero R, Hurtado M, Briega A, Parra R. Comparison of three microtube column agglutination systems for antibody screening: DG Gel, DiaMed-ID and Ortho BioVue. Transfus Med. 2006; 16:131–136.
Article
12. Han KS, Cho HI, Kim SI. A Study on the hemolytic transfusion reactions due to irregular antibodies. Korean J Hematol. 1989; 24:27–33.
13. Höglund P, Rosengren-Lindquist R, Wikman AT. A severe haemolytic transfusion reaction caused by anti-Lea active at 37℃. Blood Transfus. 2013; 11:456–459.
14. Lee DD, Oh SH, Shin HJ, Chung JS, Cho GJ, Park TS, et al. Diagnostic usefulness of antibody identification test including enzyme method. Korean J Hematol. 2006; 41:167–171.
Article
15. Weisbach V, Ziener A, Zimmermann R, Glaser A, Zingsem J, Eckstein R. Comparison of the performance of four microtube column agglutination systems in the detection of red cell alloantibodies. Transfusion. 1999; 39:1045–1050.
Article
16. Weisbach V, Kohnhäuser T, Zimmermann R, Ringwald J, Strasser E, Zingsem J, et al. Comparison of the performance of microtube column systems and solid-phase systems and the tube low-ionic-strength solution additive indirect antiglobulin test in the detection of red cell alloantibodies. Transfus Med. 2006; 16:276–284.
Article
17. Kim SH, Nam DH, Yang JH, Yoon SY, Kim YK, Lee KN, et al. Evaluation of the automatic blood bank instrument AutoVue Innova for antibody screening. Korean J Blood Transfus. 2008; 19:140–145.
18. Lim G, Park KS, Park TS, Lee HJ, Suh JT, Park SY. The frequency and distribution of unexpected antibodies in transfusion candidates with the use of the Ortho Biovue System: recent four year experience. Korean J Blood Transfus. 2009; 20:23–31.
19. Joo SY, Han KS, Kwak YK, Park KU. Evaluation of an automated solid-phase cell adherence assay in the Galileo System (Immucor) for routine pretransfusion tests. Korean J Blood Transfus. 2011; 22:134–143.
20. An TK, Song YK, Seo HS, Kim KL, Kim JA, Ko CH, et al. Evaluation of automated blood bank systems AutoVue Innova and QWALYS-3 for ABO-RhD grouping and antibody screening. Korean J Blood Transfus. 2012; 23:204–209.
21. Jang MA, Oh JW, Lee ST, Seo JY, Kim DW. Evaluation of the automated blood bank systems Galileo NEO and QWALYS-3 for ABO-RhD typing and antibody screening. Korean J Blood Transfus. 2014; 25:235–242.
22. Ko YH, Kim JM, Koo SH, Lim J, Park YC, Kwon KC. Evaluation of the automated QWALYS-3 System for ABO and RhD grouping and unexpected antibody screening. Korean J Blood Transfus. 2011; 22:144–150.
Full Text Links
  • LMO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr