Yonsei Med J.  2014 Mar;55(2):482-486.

Estimation of Body Fluid Volume by Bioimpedance Spectroscopy in Patients with Hyponatremia

Affiliations
  • 1Department of Nephrology, Yonsei University Wonju College of Medicine, Wonju, Korea. kidney74@yonsei.ac.kr

Abstract

PURPOSE
Estimation of body fluid volume in hyponatremia is useful for diagnosis and therapeutic decision-making. Physical examination has been generally used to estimate body fluid volume, but it depends on the physician's abilities. Bioimpedance spectroscopy has been suggested to be a reliable method for the estimation of body fluid volume. Therefore, this study investigated whether bioimpedance spectroscopy could replace physical examination in hyponatremia.
MATERIALS AND METHODS
The study included 30 patients with hyponatremia. At the time of the initial visit, body fluid volume was estimated simultaneously by both physical examination and bioimpedance spectroscopy. Estimation of body fluid status by clinical diagnosis was performed as well, which determined body fluid status corresponds with the most likely cause of hyponatremia (clinical body fluid estimation).
RESULTS
The results of body fluid volume estimated by physical examination, bioimpedance spectroscopy, and clinical body fluid estimation showed that 9, 10, and 9 patients, respectively, were hypervolemic; 13, 15 and 16 patients, respectively, were euvolemic; and 8, 5, and 5 patients, respectively, were hypovolemic. Cohen's kappa analysis showed a significant agreement between physical examination and bioimpedance spectroscopy (kappa coefficient, 0.632, p<0.001). In addition, bioimpedance spectroscopy showed a higher level of agreement with clinical body fluid estimation than physical examination (kappa coefficient, 0.602 vs. 0.524).
CONCLUSION
This study suggests that bioimpedance spectroscopy could replace physical examination for estimating body fluid status in hyponatremia. In addition, bioimpedance spectroscopy might correspond better with clinical diagnosis than physical examination in the estimation of body fluid status in hyponatremia.

Keyword

Hyponatremia; body fluids

MeSH Terms

Body Fluids*
Diagnosis
Humans
Hyponatremia*
Hypovolemia
Methods
Physical Examination
Spectrum Analysis*

Reference

1. Lindner G, Schwarz C. An update on the current management of hyponatremia. Minerva Med. 2012; 103:279–291.
2. López-Gómez JM. Evolution and applications of bioimpedance in managing chronic kidney disease. Nefrologia. 2011; 31:630–634.
3. Bunnag S, Pattanasombatsakul K. N-terminal-pro-brain natriuretic peptide for the differential diagnosis of hypovolemia vs. euvolemia in hyponatremic patients. J Med Assoc Thai. 2012; 95:Suppl 3. S69–S74.
4. Kyle UG, Bosaeus I, De Lorenzo AD, Deurenberg P, Elia M, Manuel Gómez J, et al. Bioelectrical impedance analysis-part II: utilization in clinical practice. Clin Nutr. 2004; 23:1430–1453.
Article
5. Aspromonte N, Cruz DN, Ronco C, Valle R. Role of bioimpedance vectorial analysis in cardio-renal syndromes. Semin Nephrol. 2012; 32:93–99.
Article
6. Machek P, Jirka T, Moissl U, Chamney P, Wabel P. Guided optimization of fluid status in haemodialysis patients. Nephrol Dial Transplant. 2010; 25:538–544.
Article
7. Cox-Reijven PL, Kooman JP, Soeters PB, van der Sande FM, Leunissen KM. Role of bioimpedance spectroscopy in assessment of body water compartments in hemodialysis patients. Am J Kidney Dis. 2001; 38:832–838.
Article
8. Wizemann V, Rode C, Wabel P. Whole-body spectroscopy (BCM) in the assessment of normovolemia in hemodialysis patients. Contrib Nephrol. 2008; 161:115–118.
Article
9. Wabel P, Chamney P, Moissl U, Jirka T. Importance of whole-body bioimpedance spectroscopy for the management of fluid balance. Blood Purif. 2009; 27:75–80.
Article
10. van den Ham EC, Kooman JP, Christiaans MH, Nieman FH, Van Kreel BK, Heidendal GA, et al. Body composition in renal transplant patients: bioimpedance analysis compared to isotope dilution, dual energy X-ray absorptiometry, and anthropometry. J Am Soc Nephrol. 1999; 10:1067–1079.
11. Moissl UM, Wabel P, Chamney PW, Bosaeus I, Levin NW, Bosy-Westphal A, et al. Body fluid volume determination via body composition spectroscopy in health and disease. Physiol Meas. 2006; 27:921–933.
Article
12. Chamney PW, Wabel P, Moissl UM, Müller MJ, Bosy-Westphal A, Korth O, et al. A whole-body model to distinguish excess fluid from the hydration of major body tissues. Am J Clin Nutr. 2007; 85:80–89.
Article
Full Text Links
  • YMJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr