J Korean Radiol Soc.  1993 Apr;29(2):332-338.

Teleradiology(TELEACE) system: Results of field trial

Abstract

We report the results of field operation of TELEACE system between Kyung-Pook National University Hospital and Ul-Jin Goon Health Care Medical Center from December, 1990 to September, 1991, which had been operated as a kind of Integrated Services Digital Nework projects by KOREA TELECOMMUNICATION Inc. Ul-Jin Goon Health Care Medical Center transmitted 414 plain radiographs to our hospital in speed of 9600BPS. Each image was composed of 1024X1024 pixelsX8 bits/pixel. In our hospital, the image files were displayed on high resolution monitor (1280×1024 pixets). Text files of image interpretations were transmitted to the health care medical center. The two radiologists who had interpreted the transmitted images, went to the health care medical center and read radiographic film with blind test method. We obtained the following results: false negative rate of 6.3%, false postitve rate of 2.4%, mean sensitivity of 81.4%, mean specificity of 96.3%, and mean accuracy of 91.3%. In predictive value of 0.05, there was no significant difference between results of these two types of radiographs. In conclusion, TELEACE system was valuable to the clinicians isolated from services of radiologists.


MeSH Terms

Delivery of Health Care
Korea
Methods
Sensitivity and Specificity
Telecommunications
X-Ray Film
Full Text Links
  • JKRS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr