J Korean Ophthalmol Soc.  2014 Jun;55(6):826-832.

Accuracy of Intraocular Lens Power Calculations According to Corneal Curvature in Short Eyes

Affiliations
  • 1Department of Ophthalmology, Saevit Eye Hospital, Goyang, Korea. zenith716@gmail.com

Abstract

PURPOSE
To evaluate the accuracy of intraocular lens (IOL) power calculations between Hoffer Q and other formulas according to corneal curvature by comparing the results of cataract surgery and calculation of chosen formulas in short axial lengths.
METHODS
We performed a retrospective analysis of patients who underwent cataract surgery from January 1st, 2012 to June 12th, 2012. The patients were selected if their axial length was below 23.00 mm and 77 patients (90 eyes) were included in the present study. The patients were divided into 2 groups according to mean corneal curvature below 44.0 D and over 45.0 D. IOL power was calculated using the Hoffer Q and SRK II, SRK-T and Holladay I formulas and the error between the calculations and refractive outcome of cataract surgery were measured. The accuracy of each formula was evaluated by comparing the error between the 2 groups.
RESULTS
Hoffer Q formula showed a higher predictive accuracy than other formulas regardless of corneal curvature in eyes with short axial lengths (p < 0.001, p = 0.023). Particularly, SRK II, SRK-T and Holladay I showed a lower predictive accuracy in eyes with flat corneal curvature than Hoffer Q (p < 0.001, p = 0.215).
CONCLUSIONS
In eyes with short axial lengths, preoperative predicted IOL power calculations showed better accuracy with Hoffer Q formula than SRK II, SRK-T and Holladay I formulas. SRK II, SRK-T and Holladay I formulas showed a lower predictive accuracy in flat corneal curvature eyes than steep corneal curvature eyes with short axial lengths. We hypothesize that SRK II, SRK-T and Holladay I tend to underestimate effective lens position in eyes with short axial lengths indicating Hoffer Q formula is more accurate.

Keyword

Corneal curvature; Formula; Hoffer Q; IOL calculation; Short eye

MeSH Terms

Cataract
Humans
Lenses, Intraocular*
Retrospective Studies

Reference

References

1. Holladay JT, Prager TC, Ruiz RS, et al. Improving the predict-ability of intraocular lens power calculations. Arch Ophthalmol. 1986; 104:539–41.
Article
2. Drexler W, Findl O, Menapace R, et al. Partial coherence inter-ferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol. 1998; 126:524–34.
Article
3. Holladay I JT, Gills JP, Leidlin J, Cherchio M. Achieving emme-tropia in extremely short eyes with two piggyback posterior cham-ber intraocular lenses. Ophthalmology. 1996; 103:1118–23.
Article
4. Olsen T, Corydon L, Gimbel H. Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg. 1995; 21:313–9.
Article
5. Drews RC. Reliability of lens implant power formulas in hyper-opes and myopes. Ophthalmic Surg. 1988; 19:11–5.
Article
6. Lee YE, Choi KR, Jun RM. Accuracy of intraocular lens power calculations according to the formulas and anterior chamber depth in short eyes. J Korean Ophthalmol Soc. 2010; 51:1338–44.
Article
7. Shin JA, Hwang KY, Kim MS. Refractive error according to the anterior chamber depth and corneal refractive power in short eyes. J Korean Ophthalmol Soc. 2013; 54:65–71.
8. MacLaren RE, Sagoo MS, Restori M, Allan BD. Biometry accu-racy using zero- and negative-powered intraocular lenses. J Cataract Refract Surg. 2005; 31:280–90.
Article
9. Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and re-gression formulas. J Cataract Refract Surg. 1993; 19:700–12.
Article
10. Hitzenberger CK. Optical measurement of the axial eye length by laser Doppler interferometry. Invest Ophthalmol Vis Sci. 1991; 32:616–24.
11. Drexler W, Findl O, Menapace R, et al. Dual beam optical coher-ence tomography: signal identification for ophthalmologic diagnosis. J Biomed Opt. 1998; 3:55–65.
Article
12. Fedorov SN, Kolinko AI. A method of calculating the optical pow-er of the intraocular lens. Vestn Oftalmol. 1967; 80:27–31.
13. Holladay JT, Prager TC, Chandler TY, et al. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg. 1988; 14:17–24.
Article
14. Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990; 16:333–40.
Article
15. Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and re-gression formulas. J Cataract Refract Surg. 1993; 19:700–12.
Article
16. Lam AK, Chan R, Pang PC. The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster. Ophthalmic Physiol Opt. 2001; 21:477–83.
17. Kim HJ, Kim HJ, Joo CK. Comparison of IOL Master, A-scan and Orbscan II for measurement of axial length and anterior chamber depth. J Korean Ophthalmol Soc. 2003; 44:1519–27.
18. McEwan JR, Massengill RK, Friedel SD. Effect of keratometer and axial length measurement errors on primary implant power calculations. J Cataract Refract Surg. 1990; 16:61–70.
Article
19. Drexler W, Findl O, Menapace R, et al. Dual beam optical coher-ence tomography: signal identification for ophthalmologic diagnosis. J Biomed Opt. 1998; 3:55–65.
Article
20. Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 1992; 18:125–9.
Article
21. Haigis W. The Haigis Formula. In : Shammas HJ, editor. Intraocular lens power calculations. Thorofare, NJ: SLACK Incorporated;2004. chap. 5.
22. Olsen T, Olesen H, Thim K, Corydon L. Prediction of post-operative intraocular lens chamber depth. J Cataract Refract Surg. 1990; 16:587–90.
Article
23. Olsen T, Corydon L, Gimbel H. Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg. 1995; 21:313–9.
Article
24. Maeng HS, Ryu EH, Chung TY, Chung ES. Effect of anterior chamber depth and axial length on refractive error after intraocular lens implantation. J Korean Ophthalmol Soc. 2010; 51:195–202.
25. Lee YE, Choi KR, Jun RM. Accuracy of intraocular lens power calculations according to the formulas and anterior chamber depth in short eyes. J Korean Ophthalmol Soc. 2010; 51:1338–44.
Article
Full Text Links
  • JKOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr