J Lab Med Qual Assur.  2002 Dec;24(2):221-224.

Evaluation of the HM-JACK Automatic Analyzer for Fecal Occult Blood Test

Affiliations
  • 1Department of Laboratory Medicine, Pundang Jesaeng General Hospital, Sungnam, Korea. dhlee@dmc.or.kr
  • 2Department of Laboratory Medicine, Sungkyunkwan University School of Medicine Kangbuk Samsung Hospital, Seoul, Korea.

Abstract

BACKGROUND: Fecal occult blood test is a useful method for mass screening of colorectal cancer. HM-JACK is automatic analyzer for detecting fecal occult blood with latex agglutination methods. We evaluated its apparatus for compatibility for clinical testing.
METHODS
The linearity, precision and effect of temperature and container in stool storage were evaluated. The comparison study with OC-Hemodia and prozone phenomenon were evaluated.
RESULTS
The linearity was good(R(2)=0.997) and coeffiecinet variation(CV) of within-day precision were 4.6% and 2.6% in low concentration and high concentration. The CV of between-day precision were 4.8% and 3.5%. The hemoglobin concentration of 0.1 g/dL(1x10(6) ng/mL) was measured to 651 ng/mL for Antigen excess zone. The hemoglobin were more decreased in room temperature storage and usual container storage than in low temperature storage and HM-JACK container storage. The concordance rate between HM-JACK and OC-Hemodia was 90%.
CONCLUSION
HM-JACK showed good performance for linearity, precision and comparison study. Therefore, it can be used effectively in clinical laboratory due to convenience, and possibility of quantitation and fast reporting.

Keyword

Fecal occult blood test; Colorectal cancer; HM-JACK; OC-Hemodia

MeSH Terms

Agglutination
Colorectal Neoplasms
Latex
Mass Screening
Occult Blood*
Latex
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr