Gut Liver.  2010 Jun;4(2):192-200.

A Pilot Study of Sequential Capsule Endoscopy Using MiroCam and PillCam SB Devices with Different Transmission Technologies

Affiliations
  • 1Division of Gastroenterology, Department of Internal Medicine and Yonsei Institute of Gastroenterology, Yonsei University College of Medicine, Seoul, Korea. sysong@yuhs.ac
  • 2Brain Korea 21 Project for Medical Science, Yonsei University College of Medicine, Seoul, Korea.

Abstract

BACKGROUND/AIMS
Studies have investigated the use of different types of radiofrequency capsules for comparison or sequential capsule endoscopy, but none have compared the MiroCam device - which utilizes a novel data transmission technology - with other capsules. This study compared the feasibility of sequential capsule endoscopy using the MiroCam and PillCam SB devices, which employ different transmission technologies.
METHODS
Patients with diseases requiring capsule endoscopy were enrolled. After a 12-hour fast, one randomly selected capsule was swallowed. The second capsule was swallowed once fluoroscopy had indicated that the first capsule had migrated below the gastric outlet.
RESULTS
The total operating time in 24 patients was 702+/-60 min (mean+/-SD) for the MiroCam and 446+/-28 min for the PillCam SB (p<0.0001). The rate of a complete examination to the cecum was 83.3% for the MiroCam and 58.3% for the PillCam SB (p=0.031). Diagnostic yields for the MiroCam, PillCam SB, and sequential capsule endoscopy were 45.8%, 41.7%, and 50.0%, respectively. The agreement rate between the two capsules was 87.5%, with a kappa value of 0.74. Electrical interference in data transmission between the two capsules was not observed, but temporary visual interferences were observed in seven patients (29.2%).
CONCLUSIONS
Sequential capsule endoscopy with the MiroCam and PillCam SB produced slight but nonsignificant increases in the diagnostic yield, and the two capsules did not exhibit electrical interference. A larger trial is necessary for elucidating the usefulness of sequential capsule endoscopy.

Keyword

Capsule endoscopy; Gastrointestinal hemorrhage; Feasibility study; Diagnosis

MeSH Terms

Capsule Endoscopy
Capsules
Cecum
Feasibility Studies
Fluoroscopy
Gastrointestinal Hemorrhage
Humans
Pilot Projects
Capsules