Ann Dermatol.  2013 Feb;25(1):67-72. 10.5021/ad.2013.25.1.67.

Reliability on Intra-Laboratory and Inter-Laboratory Data of Hair Mineral Analysis Comparing with Blood Analysis

Affiliations
  • 1Department of Dermatology, Medical College, Dankook University, Cheonan, Korea. 4exodus@hanmail.net

Abstract

BACKGROUND
Nowadays, although its clinical value remains controversial institutions utilize hair mineral analysis. Arguments about the reliability of hair mineral analysis persist, and there have been evaluations of commercial laboratories performing hair mineral analysis.
OBJECTIVE
The objective of this study was to assess the reliability of intra-laboratory and inter-laboratory data at three commercial laboratories conducting hair mineral analysis, compared to serum mineral analysis.
METHODS
Two divided hair samples taken from near the scalp were submitted for analysis at the same time, to all laboratories, from one healthy volunteer. Each laboratory sent a report consisting of quantitative results and their interpretation of health implications. Differences among intra-laboratory and interlaboratory data were analyzed using SPSS version 12.0 (SPSS Inc., USA).
RESULTS
All the laboratories used identical methods for quantitative analysis, and they generated consistent numerical results according to Friedman analysis of variance. However, the normal reference ranges of each laboratory varied. As such, each laboratory interpreted the patient's health differently. On intra-laboratory data, Wilcoxon analysis suggested they generated relatively coherent data, but laboratory B could not in one element, so its reliability was doubtful. In comparison with the blood test, laboratory C generated identical results, but not laboratory A and B.
CONCLUSION
Hair mineral analysis has its limitations, considering the reliability of inter and intra laboratory analysis comparing with blood analysis. As such, clinicians should be cautious when applying hair mineral analysis as an ancillary tool. Each laboratory included in this study requires continuous refinement from now on for inducing standardized normal reference levels.

Keyword

Blood minerals; Hair minerals; Minerals; Scalp hair

MeSH Terms

Hair
Hematologic Tests
Minerals
Reference Values
Scalp
Minerals

Cited by  1 articles

Hair Mineral Analysis in Children with Atopic Dermatitis
Jeong Eun Kim, Jae Min Shin, Joo Yeon Ko, Young Suck Ro
Ann Dermatol. 2017;29(2):251-253.    doi: 10.5021/ad.2017.29.2.251.


Reference

1. Bass DA, Hickock D, Quig D, Urek K. Trace element analysis in hair: factors determining accuracy, precision, and reliability. Altern Med Rev. 2001. 6:472–481.
2. Suliburska J, Bogdański P, Pupek-Musialik D, Krejpcio Z. Dietary intake and serum and hair concentrations of minerals and their relationship with serum lipids and glucose levels in hypertensive and obese patients with insulin resistance. Biol Trace Elem Res. 2011. 139:137–150.
Article
3. Kim OY, Baek SH, Lee YJ, Lee KH. Association of increased hair calcium levels and enhanced augmentation index (AIx): a marker of arterial stiffness. Biol Trace Elem Res. 2010. 138:90–98.
Article
4. Park SB, Choi SW, Nam AY. Hair tissue mineral analysis and metabolic syndrome. Biol Trace Elem Res. 2009. 130:218–228.
Article
5. Yasuda H, Yoshida K, Segawa M, Tokuda R, Tsutsui T, Yasuda Y, et al. Metallomics study using hair mineral analysis and multiple logistic regression analysis: relationship between cancer and minerals. Environ Health Prev Med. 2009. 14:261–266.
Article
6. Rautio A, Kunnasranta M, Valtonen A, Ikonen M, Hyvärinen H, Holopainen IJ, et al. Sex, age, and tissue specific accumulation of eight metals, arsenic, and selenium in the European hedgehog (Erinaceus europaeus). Arch Environ Contam Toxicol. 2010. 59:642–651.
Article
7. Grabeklis AR, Skalny AV, Nechiporenko SP, Lakarova EV. Indicator ability of biosubstances in monitoring the moderate occupational exposure to toxic metals. J Trace Elem Med Biol. 2011. 25:Suppl 1. S41–S44.
Article
8. Barrett S. Commercial hair analysis. Science or scam? JAMA. 1985. 254:1041–1045.
Article
9. Miekeley N, Dias Carneiro MT, da Silveira CL. How reliable are human hair reference intervals for trace elements? Sci Total Environ. 1998. 218:9–17.
Article
10. Druyan ME, Bass D, Puchyr R, Urek K, Quig D, Harmon E, et al. Determination of reference ranges for elements in human scalp hair. Biol Trace Elem Res. 1998. 62:183–197.
Article
11. Seidel S, Kreutzer R, Smith D, McNeel S, Gilliss D. Assessment of commercial laboratories performing hair mineral analysis. JAMA. 2001. 285:67–72.
Article
12. Drasch G, Roider G. Assessment of hair mineral analysis commercially offered in Germany. J Trace Elem Med Biol. 2002. 16:27–31.
Article
13. Steindel SJ, Howanitz PJ. The uncertainty of hair analysis for trace metals. JAMA. 2001. 285:83–85.
Article
14. Hamilton T, Schweinsberg F. The reliability of conclusions based on hair mineral analysis in individual diagnostic. Versicherungsmedizin. 2004. 56:136–140.
Full Text Links
  • AD
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr