Korean J Nosocomial Infect Control.
2000 Jun;5(1):1-8.
Comparison of Mycobactericidal Activity of 12 Kinds of Disinfectants for Mycobacterium chelonae
- Affiliations
-
- 1Department of Clinical Pathology, College of Medicine, The Catholic University of Korea, Seoul, Korea.
- 2Department of Hospital Infection Control of Kangnam St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea.
Abstract
- BACKGROUND
Cleaning and disinfection of fiberoptic bronchoscope requires careful attention, especially to mycobacterium species because the contamination of mycobacteria could raise confusion on diagnosis. Recently, we detected contamination of Wydex(R) solution used in bronchoscope washer with Mycobacterium chelonae. In this study, we evaluated the mycobactericidal effect of 12 kinds of disinfectants for M. chelonae.
METHOD: To evaluate the bactericidal effect of Wydex(R) 2%, Cidex(R) 2.25%, Cidex(R) 3%, Bacteriokiller (BK) disinfectant, Perasafe(R), HICLO-S(R), Lamicine(R), ethanol, Instrusept(R), Virkon(R), Betadine(R), and Vipon(R) against M. chelonae, culture was performed after exposure of two M. chelonae strains (ATCC 35749, the type strain and the strain isolated from contaminated Wydex(R) solution) to each disinfectant solution. The growth of organism was examined for up to 8 weeks.
RESULTS
Growth of M. chelonae (reference strain of ATCC 35749 and isolated strain) was observed after a week incubation for Wydex(R) 2%, Cidex(R) (2.25%, 3%) and control. For BK disinfectant and Perasafe(R), they grew after 2-3 weeks, and 3-4 weeks, respectively. For HICLO-S(R) and Lamicine(R), only the contaminated strain grew after two and three weeks, respectively. For ethanol, Virkon(R), Betadine(R), Vipon(R), and Instrusept(R) , growth was not observed from either strain.
CONCLUSIONS
On the basis of these results, Instrusept(R), virkon(R), ethanol, Betadine(R), and Vipon(R) were effective for the disinfection of M. chelonae. Especially, Instrusept(R) was thought to be useful as a disinfectant for bronchoscopes because it has advantages including non-corrosiveness, chemical stability, and non-irritativeness. And additional washing with ethanol might be effective. The finding that strain isolated from contaminated bronchoscopes was more resistant to disinfectants than reference strain suggested that the more resistant strains are selected throughout the improper disinfection.