Effect of Transverse Magnetic Field on Dose Distribution of High Energy Electron Beam
- Affiliations
-
- 1Department of Radiation Oncology, Eulji University Hospital, Daejeon, Korea.
- 2Department of Radiation Oncology, Chungnam National University Hospital, Daejeon, Korea. mjcho@cnu.ac.kr
- 3Department of Radiation Oncology, Dankook University Hospital, Cheonan, Korea.
- 4Research Institute of Clinical Medicine, Chonbuk National University, Jeonju, Korea.
- 5Department of Radiation Oncology, Dong-A University Hospital, Busan, Korea.
- 6Department of Radiation Oncology, School of Medicine, Wonkwang University, Iksan, Korea.
- 7Cancer Research Institute, Chungnam National University, Daejeon, Korea.
- 8Department of Radiation Oncology, Yeongnam University Hospital, Daegu, Korea.
Abstract
-
In this work we have measured the dose distribution and the percent depth dose of 20 MeV electron beam using the X-OMAT films in order to verify the effects of transverse magnetic field on high energy elecrtron beam in a phantom. The result shows about 30% increase of the percent depth dose at 4.5 cm depth under the transverse magnetic field of 1.5 Tesla at 7.5 cm depth. We have verified that these were in an agreement with other theoretical results.