Korean J Clin Pathol.  1997 Apr;17(2):218-229.

Comparison Study of Reticulocyte Enumeration by H*3 RTX(TM), FACScan(TM), and Manual Counting

Abstract

BACKGROUND: Reticulocyte counts provide clinically useful informations and the most widely used method for counting reticulocytes is a manual microscopic procedure. Although manual method is inexpensive and relatively simple to perform, it is labor intensive and imprecise. So, more rapid and more reproducible methods are needed.
METHODS
Reticulocyte counts on 96 blood samples were performed by conventional manual method, H*3 RTX(TM)(Bayer, U.S.A.), and FACScan(TM)(Beckon Dickinson, U.S.A.). The changes of reticulocyte count after storage of samples and certain time intervals after preparation were also examined.
RESULTS
Reticulocyte counts(%) by manual method, H*3 RTX(TM), and FACScanM were 2.05+/-2.16, 1.95+/-2.24 and 2.51+/-1.94, respectively. There was no statistically significant differences between manual counting and H*3 RTX(TM)(P>0.05). However, there were significant differences between H*3 RTX(TM) and FACScan(TM), manual counting and FACScan(TM)(P<0.05). Correlation coefficients of three comparisons were all above 0.920. Statistically not confirmed the reticulocytosis sample showed decreasing tendency of reticuocyte count by H*3 RTX(TM) after storage of the sample and prepared sample and others were relatively stable in H*3 RTX(TM)and FACScan(TM).
CONCLUSIONS
The result of the mean difference between H*3 RTX(TM) and manual counting is statistically insignificant. So, the H*3 RTX(TM)can be used interchangeably with manual counting atter consideration of cost-effectiveness.


MeSH Terms

Reticulocyte Count
Reticulocytes*
Reticulocytosis