Electromagnetic Interference of Wireless Local Area Network on Electrocardiogram Monitoring System: A Case Report

Affiliations
  • 1Division of Cardiology, Cardiac and Vascular Center, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. s.woo.park@samsung.com
  • 2Department of Information Strategy, Samsung Medical Center, Seoul, Korea.

Abstract

Electromagnetic interference (EMI) can affect various medical devices. Herein, we report the case of EMI from wireless local area network (WLAN) on an electrocardiogram (ECG) monitoring system. A patient who had a prior myocardial infarction participated in the cardiac rehabilitation program in the sports medicine center of our hospital under the wireless ECG monitoring system. After WLAN was installed, wireless ECG monitoring system failed to show a proper ECG signal. ECG signal was distorted when WLAN was turned on, but it was normalized after turning off the WLAN.

Keyword

Wireless technology; Electrocardiography

MeSH Terms

Electrocardiography
Humans
Local Area Networks
Magnets
Myocardial Infarction
Sports Medicine
Wireless Technology

Figure

  • Fig. 1 Electromagnetic interference of wireless local area network (WLAN) on ECG monitor. A distorted ECG signal was displayed on the ECG monitor after placing ECG leads when the WLAN was turned on (A). The ECG was normalized after turning off WLAN (B). The abnormal ECG pattern reappeared right after turning on the WLAN (C). ECG: electrocardiogram.


Reference

1. Gladman AS, Lapinsky SE. Wireless technology in the ICU: boon or ban? Crit Care. 2007. 11:165.
2. Lapinsky SE, Easty AC. Electromagnetic interference in critical care. J Crit Care. 2006. 21:267–270.
3. Klein AA, Djaiani GN. Mobile phones in the hospital--past, present and future. Anaesthesia. 2003. 58:353–357.
4. Tri JL, Severson RP, Firl AR, Hayes DL, Abenstein JP. Cellular telephone interference with medical equipment. Mayo Clin Proc. 2005. 80:1286–1290.
5. Hekmat K, Salemink B, Lauterbach G, et al. Interference by cellular phones with permanent implanted pacemakers: an update. Europace. 2004. 6:363–369.
6. Fetter JG, Ivans V, Benditt DG, Collins J. Digital cellular telephone interaction with implantable cardioverter-defibrillators. J Am Coll Cardiol. 1998. 31:623–628.
7. Shaw CI, Kacmarek RM, Hampton RL, et al. Cellular phone interference with the operation of mechanical ventilators. Crit Care Med. 2004. 32:928–931.
8. Baranchuk A, Kang J, Shaw C, et al. Electromagnetic interference of communication devices on ECG machines. Clin Cardiol. 2009. 32:588–592.
9. Kapa S, Pierce T, Hayes DL, Holmes DR Jr, Asirvatham SJ. Electromagnetic interference of magnetic field based auto identification technologies in healthcare settings. Int J Med Inform. 2011. 80:239–250.
10. García CU, Sigler FG, Duran MD, et al. On practical issues about interference in telecare applications based on different wireless technologies. Telemed J E Health. 2007. 13:519–533.
Full Text Links
  • KCJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr