J Korean Ophthalmol Soc.  2016 Jan;57(1):20-24. 10.3341/jkos.2016.57.1.20.

Evaluation of Objective Accommodation Power in Different Age Groups Using an Auto Accommodation Refractometer

Affiliations
  • 1Department of Ophthalmology, Korea University College of Medicine, Seoul, Korea. crisim@korea.ac.kr

Abstract

PURPOSE
To evaluate the objective amplitude of accommodation in various age groups using an autorefractometer and to compare the results with subjective assessments on accommodation. Objective accommodative amplitude of artificial lenses in post-operative cataract patients was also measured.
METHODS
In this study, 41 patients who visited Korea University Guro Hospital from July through August 2014 were categorized into 6 different age groups. Accommodative amplitude was measured using long-distance refraction at 5 m and short-distance refraction at 20 cm. In 20 patients, subjective amplitude of accommodation was measured using the minus lens technique and was compared with the measured objective accommodative amplitude. Accommodative amplitude was also measured in 8 patients who received cataract surgery.
RESULTS
The mean accommodative ability in normal people was 2.38 D. The age groups of 10-19, 20-29, 30-39, 40-49, 50-59, and over 60 years showed accommodative amplitudes of 3.31 +/- 0.99 D, 3.25 +/- 0.34 D, 3.78 +/- 0.68 D, 1.00 +/- 0.60 D, 0.22 +/- 0.23 D, 0.00 +/- 0.00 D, respectively. The results showed a rapid decrease in accommodative ability of patients older than 40 years and demonstrated a correlation between objective and subjective amplitude of accommodation (Pearson's correlation coefficient 0.838). There was a significant difference between objective and subjective accommodative amplitudes in artificial lenses (p = 0.015), with average objective and subjective measures of 0.10 +/- 0.23 D and 0.59 +/- 0.35 D, respectively.
CONCLUSIONS
Measuring amplitude of accommodation using an autorefractometer can be useful in assessing results of presbyopia treatment.

Keyword

Accommodation; Autorefractometer; Pseudoaccommodation

MeSH Terms

Cataract
Humans
Korea
Presbyopia

Figure

  • Figure 1. Measurement of objective accommodation power us-ing autorefractometer (WAM-5500, Grand Seiko, Hiroshima, Japan).

  • Figure 2. Comparison of mean accommodation power be-tween subjective and objective measurements according to dif-ferent age groups.

  • Figure 3. Correlation between objective and subjective ac-commodation power in phakic eye.

  • Figure 4. Correlation between objective and subjective ac-commodation power in pseudophakic eye.


Cited by  1 articles

Relationship between Ocular Fatigue and Use of a Virtual Reality Device
Sang Hyeok Lee, Martha Kim, Hyosun Kim, Choul Yong Park
J Korean Ophthalmol Soc. 2020;61(2):125-137.    doi: 10.3341/jkos.2020.61.2.125.


Reference

References

1. Michaels DD. Visual Optics and Refraction. 1st. St. Louis: The C.V;Mosby company. 1975. p. 268–81.
Article
2. Bettman JW. Apparent accommodation in aphakic eyes. Am J Ophthalmol. 1950; 33:921–8.
Article
3. Sugitami Y, Komori T, Kitoh R, Hayano S. Apparent accom-modation (pseudo-accommodation) on pseudophakia. Folia Ophthalmol Jpn. 1979; 30:326–31.
4. Huber C. Planned myopic astigmatism as a substitute for accom-modation in pseudophakia. J Am Intraocul Implant Soc. 1981; 7:244–9.
Article
5. Nakazawa M, Ohtsuki K. Apparent accommodation in pseudo-phakic eyes after implantation of posterior chamber intraocular lenses: optical analysis. Invest Ophthalmol Vis Sci. 1984; 25:1458–60.
6. Ravalico G, Baccara F. Apparent accommodation in pseudophakic eyes. Acta Ophthalmol (Copenh). 1990; 68:604–6.
Article
7. Duane A. Are the current theories of accommodation correct? Am J Ophthalmol. 1925; 8:196–202.
8. Kabayama T, Fuseya Y, Miyazaki H. A quasi-static study of ac-commodation with aging in normal cases. Nihon Ganka Gakkai Zasshi. 1987; 91:494–7.
9. Kim BC. The normal value of amplitude of accommodation in Koreans. J Korean Ophthalmol Soc. 1979; 20:153–7.
10. Sheppard AL, Davies LN. Clinical evaluation of the Grand Seiko Auto Ref/Keratometer WAM-5500. Ophthalmic Physiol Opt. 2010; 30:143–51.
Article
11. Win-Hall DM, Glasser A. Objective accommodation measurements in prepresbyopic eyes using an autorefractor and an aberrometer. J Cataract Refract Surg. 2008; 34:774–84.
Article
12. Win-Hall DM, Glasser A. Objective accommodation measure-ments in pseudophakic subjects using an autorefractor and an aberrometer. J Cataract Refract Surg. 2009; 35:282–90.
Article
13. Harvey EM, Miller JM, Apple HP. . Accommodation in astig-matic children during visual task performance. Invest Ophthalmol Vis Sci. 2014; 55:5420–30.
Article
14. Zamora-Alejo KV, Moore SP, Parker DG. . Objective accom-modation measurement of the Crystalens HD compared to mono-focal intraocular lenses. J Refract Surg. 2013; 29:133–9.
Article
Full Text Links
  • JKOS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr