BACKGROUND
A handheld blood gas analyser has been newly developed for mobile monitoring of blood gasses and electrolytes. We evaluated the performance of a portable blood gas analyser, Epoc(TM) (Epocal Inc., Canada) according to Clinical and Laboratory Standard Institute (CLSI) guidelines, and compared it to that of a conventional analyser used in clinical laboratories.
METHODS
Precision and percent carry-over were determined using three levels of aqueous and hematocrit control materials according to CLSI EP10-A3. Linearity was determined using five levels of control materials according to CLSI EP6-A. The pH, pCO2, pO2, Na+, K+, Ca2+, glucose, lactate, and hematocrit levels were compared between the Epoc(TM) and Stat Profile Critical Care Xpress (STP CCX; Nova Biomedical, USA) analysers using whole blood samples from 40 subjects according to CLSI EP9-A2.
RESULTS
The coefficient of variation of the within-run precision and total precision were 0.000% to 4.563% and 0.000% to 5.298%, respectively. The carry-over was within 5%. The Epoc(TM) analyser showed excellent linearity for all nine parameters evaluated. For the comparison study, the Epoc(TM) and conventional analysers showed comparable results (correlation coefficient [r]=0.900-0.995), except for hematocrit (r=0.764).
CONCLUSIONS
The Epoc(TM) POC analyser shows reliable analytical precision and is comparable to the traditional bench-top blood gas analysers. It could be useful in clinical settings, especially in operating rooms and pediatric intensive care units.

Keyword

Analytical chemistry; Blood gas analysis; Point-of-care system

MeSH Terms

Blood Gas Analysis
Critical Care
Electrolytes
Glucose
Hematocrit
Hydrogen-Ion Concentration
Intensive Care Units, Pediatric
Lactic Acid
Operating Rooms
Point-of-Care Systems
Electrolytes
Glucose
Lactic Acid
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr