J Korean Orthop Assoc.  2001 Dec;36(6):623-629.

Clinical Application of Unilateral-Ring Hybrid Fixation Using DynaExtor(R)

Affiliations
  • 1Department of Orthopedic Surgery, College of Medicine, Seoul National University, Seoul, Korea.

Abstract

PURPOSE
We have used a hybrid fixation, in which DynaExtor(R) and Ilizarov rings are connected each other with special connectors. Here, we report upon the technical tips learned and the clinical outcomes of the first twenty five consecutive cases treated using this method.
MATERIALS AND METHODS
Between May 1997 and December 1999, 25 patients were treated using hybrid fixation aimed at simple lengthening (8 patients), concomitant deformity correction and lengthening (12), deformity correction alone (3), and fracture fixation (2).
RESULTS
In the group requiring simple lengthening, the healing index (H.I.) was 38.3 day/cm (23.6-66.8 day/cm). For concomitant deformity correction and lengthening, the mean amount of angular correction was 33.8degrees (15-75degrees) and the length gain, 4.7 cm (2.2-8.9 cm). For deformity correction alone, the amount of angular correction was 50.1degrees (15-120degrees). For fracture fixation, the mean external fixation time was 59.5 days (26-93 days). Complications included mild pin-tract infection in 8 patients, a temporary stiff joint in 5, and equinus of the foot and delayed union in one patient each.
CONCLUSION
The DynaExtor(R) hybrid-fixation system is an effective and convenient method that can be selectively applied to cases that need deformity correction, lengthening, or fixation of the long bone.

Keyword

Unilateral-ring hybrid fixation; Bone lengthening; Deformity correction

MeSH Terms

Bone Lengthening
Congenital Abnormalities
Foot
Fracture Fixation
Humans
Joints
Full Text Links
  • JKOA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr