J Korean Knee Soc.  1999 Dec;11(2):231-235.

Breakage of the Implant after Total Knee Replacement: Cases Report

Affiliations
  • 1Department of Orthopaedics Surgery, Wonkwang University Hospital, Iksan, Korea.

Abstract

After total knee replacement, the breakage of implant is unusual. We have been experienced the two cases, one is the porous tibial plate breakage that Miller/Galante total knee system(MG TKS) and the other is the metal-backed polyethylene patellar component breakage with low contact stress(LCS) pros-thesis. Revisions have been performed by using Nexgen and modular posterior stabilized P.F.C. total knee system. Considerations of these causes are habitual abrupt hyperflexion of the knee and weight bear-ing. So we need to careful observation of long term clinical, radiological and survival results of total knee prosthesis in regard of the breakage of implant.

Keyword

LCS; Patellar bearing; MG Porous tibial plate; Breakage