J Korean Ophthalmol Soc.
2000 Jan;41(1):172-178.
A Comparison of Frequency Doubling Technology N-30 and Humphrey Field Analyzer Central 30-2
- Affiliations
-
- 1The Institute of Vision Research, College of Medicine, Yonsei University, Seoul, Korea.
- 2Department of Ophthalmology, Yonsei University College of Medicine.
Abstract
- To evaluate the validity of FDT as a field analyzer by comparing to Humphrey perimetry with global indeces.FDT is known as a easy to use and convenient device, and there is restriction in pupil size and refractive error within 7 diopters. Forty-two glaucoma patients[79 eyes]were included in this study. All subjects performed FDT N-30 and HFA C30-2 apart 15 minutes on same day.Mean age of subjects were 48.18 +/-16.46years[from sixteen to seventy four years].Twenty-five men and seventeen women were included in this study. Test duration was 4.92 +/-1.08 minutes with FDT N-30 and 13.79 +/-3.32 minutes with HFA C30-2.In global index mean deviation was 14.63 +/-9.52 decibel with HFA C30-2 and 10.33 +/-6 .8 3 decibel with FDT.There was a significant correlation with coefficient correlation 0.940 between two tests[p<0.01]. Pattern standard deviation was 7.89 +/-3 .8 5 decibel in HFA C30-2 and 7.80 +/-3.13 in FDT.There was a significant correlation with coefficient correlation 0.702 between FDT PSD and HFA C30-2 PSD, and 0.691 between FDT PSD and HFA C30-2 CPSD, too[P<0.01].In diagnosing glaucoma, sensitivity and specificity of FDT was 86% and 70% respectively. FDT is thought to be convenient and effective screening tool in glaucoma patients for with a high correlation with HFA C30-2.