Mid-term Result of the Transcatheter Occlusion of Patent Ductus Arteriosus with Duct-Occlud Device and Procedure-Related Problems
- Affiliations
-
- 1Division of Pediatric Cardiology, Cardiovascular Surgery, Yonsei Cardiovascular Center, Korea. cjy0122@yumc.yonsei.ac.kr
- 2Cardiovascular Research Institute, Yonsei University College of Medicine, Seoul, Korea.
Abstract
- PURPOSE
We will present our mid-term result of transcatheter closure of PDA with Duct-Occlud device(pfm. AG. Germany) after 12 months follow up and report the problems during the procedure.
METHODS
In total 154 patients, the Duct-Occlud devices were inserted in our institute from March, 1996 to August, 2002. Three types of Duct-Occlud device, i.e standard, reinforced, reinforced reverse cone coil were used. Echocardiographic examination was performed at 1, 6, 12 months after procedure.
RESULTS
The echocardiographic closure rate was 96% after 12 months. The rates of residual shunt in the standard coil, the reverse cone coil, and the reinforced reverse cone group were 8%, 4% and 3% respectively. In PDA with diameter less than 4 mm, the closure rate was up to 98% while in large PDA with more than 4 mm, it was 72% after 12 months. Embolization of the inserted coils had occurred in 5 cases with successful retrieval using snare catheter. The rupture of the core wire during the procedure and distortion of the original coil shape had occurred in 4 cases.
CONCLUSION
The transcatheter occlusion with Duct Occlud is safe and effective method for small to moderate sized PDA less than 4 mm. The minimum diameter of the PDA seems to be the predictor of residual shunt. Further refinement of the device to overcome the procedure-related problems seems to be needed.