J Korean Soc Aesthetic Plast Surg.  1998 Mar;4(1):52-59.

The Reliability of Ultrasonographic Fending of Silicone Breast Implant Rupture

Abstract

This retrospective study evaluated possible rupture of silicone gel breast implants in 92 patients. the series included patients undergone augmentation mammoplasty(n=59), and reconstructive mammoplasty(n=23) from Sep. 1993 to Aug. 1996. the age of implants ranged from 4 months to 8 years(mean:23.1months). of these, 19 cases displayed Ultrasonographic sings of rupture. of 13 implants removed, 7 were intact and 6 were ruptured. Implant contour deformities and radial folds are often seen in both ruptured and intact silicone implants and therefore cannot serve as reliable signs of rupture. A stepladder sign in intact implant is believed to be the result of reverberation artifacts within the interior of the implant. Due to these false-positive ultrasonographic findings, ultrasonographiy is not as absolutely reliable tool for the diagnosis of implant rupture. Alternative imaging methods(CT, MRI) are required to establish an accurate preoperative diagnosis.


MeSH Terms

Artifacts
Breast Implants*
Breast*
Congenital Abnormalities
Diagnosis
Humans
Retrospective Studies
Rupture*
Silicone Gels*
Silicone Gels
Full Text Links
  • JKSAPS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr