J Korean Ophthalmol Soc.
1997 Apr;38(4):659-665.
An Effects of Contact Transscleral Diode Laser Cyclophotocoagulation in Glaucoma Patients
- Affiliations
-
- 1Department of Ophthalmology, College of Medicine, Dong-A University, Pusan, Korea.
Abstract
-
We observed the changes of IOP after contact transscleral cyclophotocoagualtion with Diode laser which emit 810mm beam and we tride to investigatethe suitable level and range of energy when Diode laser was applied to the refractory glaucoma patients. Fifteen eyes underwent contact transscleral cyclophotocoagulation with energy 3J(Group A, 6eyes), 4J(Group B, 9eyes) separately and intraocular pressure(IOP) was measured at 1 day, 1 week, 1 month, 2 month, 3 month, 4month, 5month, 6month, postoperatively. The intraocular pressure(IOP) decreasing rate was 83.02% in A group and 64.72% in B group. The success rate was 50% in A group and 66.7% in B group at 6 month postoperatively. It is suggested that contact transscleral cyclophotocoagulation with Diode laser si useful in lowering IOP in refractory glaucoma patients.