J Korean Orthop Assoc.  2001 Aug;36(4):373-376.

Polyethylene Liner Wear in Harris-Galante Acetabular Cup: Two Dimensional versus Three Dimensional Method

Affiliations
  • 1Department of Orthopedic Surgery, Korea University, Seoul, Korea.

Abstract

PURPOSE
The purpose of this study was to compare the two-dimensional with three-dimensional radiographic measurements of polyethylene liner wear in the Harris-Galante II acetabular cup.
MATERIALS AND METHODS
We measured the polyethylene liner wear amount and the wear rate of 64 hips with Harris-Galante II acetabular cup by two methods: Livermore's method and modified Devane's method. We evaluated the relationships between the two methods.
RESULTS
The average total amount of wear and the average wear rate are 0.86+/-1.01 mm and 0.18+/-0.20 mm/year using the two-dimensional method and 0.99+/-1.01 mm and 0.21+/-0.2 mm/year using three-dimensional method respectively. The amount of wear and the wear rate by the three-dimensional method was larger than those by the two dimensional method.
CONCLUSION
The amount of polyethylene liner wear can be measured by the two-dimensional method because it is highly correlated with the amount that is measured by the three-dimensional method, since there was no differences between the two methods.

Keyword

Total hip replacement arthroplasty; Measurement of polyethylene liner wear; Two dimensional method; Three dimensional method

MeSH Terms

Acetabulum*
Hip
Polyethylene*
Polyethylene
Full Text Links
  • JKOA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr