J Korean Orthop Assoc.  2000 Oct;35(5):765-770.

Comparison Between Standard and elevated-rim Acetabular Liner in Total Hip Arthroplasty

Affiliations
  • 1Department of Orthopedic Surgery, Kwangju Christian Hospital, Kwangju, Korea.

Abstract

PURPOSE
Elevated-rim acetabular liner enhances stability in total hip arthroplasty and has been widely accepted. But it has potential adverse effects. So, we assessed the effects of the elevated-rim acetabular liner clinically and radiologically.
MATERIALS AND METHODS
We reviewed the results of 74 total hip arthroplasties of 64 patients that had been performed from January 1993 to March 1995. Thirty-six cases of the 74 cases used elevated rim acetabular liner and 38 cases did standard liner. Dislocation rate, Harris hips score, osteolysis and revision rate were compared between two groups.
RESULTS
There were no significant differences in mean anteversion, abduction angles of the cup and femoral offsets between two groups. Four of the 38 cases with the standard liner had a dislocation, compared with one of 36 cases with the elevated-rim liner (P=0.358) . Femoral osteolysis was observed 7 cases (18.4%) in the standard liner group and 14 cases (38.8%) in the elevated-rim liner group. Acetabular osteolysis was observed 17 cases (44.7%) in the standard group and 21 cases (58.3%) in the elevated-rim group. Nine cases (23.6%) had been revised in the standard group and 6 cases (16.6%) in the elevated group.
CONCLUSION
Excessive wear and breakage of the polyethylene by impingement was developed in the elevated-rim polyethylene liner.

Keyword

Total hip arthroplasty; Acetabular polyethylene liner; Elevated-rim

MeSH Terms

Acetabulum*
Arthroplasty
Arthroplasty, Replacement, Hip*
Dislocations
Hip
Humans
Osteolysis
Polyethylene
Polyethylene
Full Text Links
  • JKOA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr