Korean J Nucl Med.
2005 Jun;39(3):191-199.
Evaluation of a Conjugate View Method for Determination of Kidney Uptake
- Affiliations
-
- 1Division of Nuclear Medicine, Yonsei University College of Medicine, Yonsei University, Korea.
- 2Research Institute of Radiological Science, Department of Radiology, Yonsei University College of Medicine, Yonsei University, Korea. hjkim@yumc.yonsei.ac.kr
- 3Department of Surgery, Yonsei University College of Medicine, Yonsei University, Korea.
Abstract
- PURPOSE
In order to obtain better quantitation of kidney uptake, this study is to evaluate a conjugate view method (CVM) using a geometric mean attenuation correction for kidney uptake and to compare it to Gate's method. MATERIALS AND METHODS: We used a Monte Carlo code, SIMIND and a Zubal phantom, to simulate kidney uptake. SIMIND was both simulated with or without scatter for the Zubal phantom. Also, a real phantom test was carried out using a dual-head gamma camera. The activity of 0.5 mCi was infused into two small cylinder phantoms of 5 cm diameter, and then, they were inserted into a cylinder phantom of 20 cm diameter. The results by the CVM method were compared with ideal data without both of attenuation and scatter and with Gate's method. The CVM was performed with or without scatter correction. The Gate's method was performed without scatter correction and it was evaluated with regards to 0.12 cm (-1) and 0.15 cm (-1) attenuation coefficients. Data were analyzed with comparisons of mean counts in the regions of interest (ROI), profiles drawn over kidney images and linear regression. Correlation coefficients were calculated with ideal data, as well. RESULTS: In the case of the computer simulation, mean counts measured from ideal data, the CVM and the Gate's method were (right: 998 +/- 209, left: 896 +/- 249), (right: 911 +/- 207, left: 815 +/- 265), and (right: 1065 +/- 267, left: 1546 +/- 267), respectively. The ideal data showed good correlation with the CVM and the correlation coefficients of the CVM, Gate's method were (right: 0.91, left: 0.93) and (right: 0.85, left: 0.90), respectively. CONCLUSION: The conjugate view method using geometric mean attenuation correction resulted in better accuracy than the Gate's method. In conclusion, the conjugate view method independent of renal depths may provide more accurate kidney uptake.