Fetal Well-being Evaluation by Determination of High Frequency Measurement Noise for Fetal Heart Rate Signal
- Affiliations
-
- 1Department of Obstetric and Gynecology, Medical College, Catholic University, Seoul, Korea. mamacare@catholic.ac.kr
- 2Department of Physics, Pohang University of Science and Technology, Pohang, Korea.
Abstract
OBJECTIVE
This study was done to propose a new method of determining the cutoff frequency against measurement noise in fetal heart rate signal and to analyze the discriminating value for normal fetus (pH>7.15) and abnormal fetus (pH< or =7.15) by nonlinear method.
METHODS
The patients were classified by umbilical artery pH (7.15) or base excess (-10 mEq/L). After removing the high frequency noise components at the cutoff frequency determined in this study, we calculated correlation dimension as a discriminating value for the normal and abnormal fetuses.
RESULTS
After low pass filtering of fetal heart rate signal at the cutoff frequency, the numerical result of embedding dimension was 5.47+/-0.67 (normal), 4.43+/-0.60 (abnormal) (p<0.001), and correlation dimension was 2.51+/-0.43 (normal), 1.79+/-0.15 (abnormal).
CONCLUSION
We could determine the cutoff frequency from the different fluctuation level between high frequency noise and principal signal according the cutoff frequency function. The detection of fetal distress was done objectively by this method. We could discriminate between normal fetus and abnormal fetus by correlation dimension as indicator for the nonlinear complexity.