Korean J Oral Maxillofac Radiol.  2010 Dec;40(4):171-178.

Surgical stent for dental implant using cone beam CT images

Affiliations
  • 1Department of Oral and Maxillofacial Radiology, School of Dentistry, Kyung Hee University, Korea. omrcys@khu.ac.kr

Abstract

PURPOSE
The purpose of this study is to develop a surgical stent for dental implant procedure that can be easily applied and affordable by using cone beam computerized tomography (CBCT).
MATERIALS AND METHODS
Aluminum, Teflon-PFA (perfluoroalkoxy), and acetal (polyoxymethylene plastic) were selected as materials for the surgical stent. Among these three materials, the appropriate material was chosen using the CBCT images. The surgical stent, which could be easily placed into an oral cavity, was designed with chosen material. CBCT images of the new surgical stent on mandible were obtained using Alphard-3030 dental CT system (Asahi Roentgen Co., Ltd., Kyoto, Japan). The point of insertion was prescribed on the surgical stent with the multiplanar reconstruction software of OnDemand3D (CyberMed Inc., Seoul, Korea). Guide holes were made at the point of insertion on the surgical stent using newly designed guide jig. CBCT scans was taken for the second time to verify the accuracy of the newly designed surgical stent.
RESULTS
Teflon-PFA showed radiologically excellent image characteristics for the surgical stent. High accuracy and reproducibility of implantation were confirmed with the surgical stent.
CONCLUSION
The newly designed surgical stent can lead to the accurate implantation and achieve the clinically predictable result.

Keyword

Cone-Beam Computed Tomography; Dental Implants; Surgical Stent

MeSH Terms

Aluminum
Cone-Beam Computed Tomography
Dental Implants
Mandible
Mouth
Stents
Aluminum
Dental Implants
Full Text Links
  • KJOMR
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr