Anesth Pain Med.  2012 Jul;7(3):221-225.

Electrical stimulation of the epiradicular space using a catheter with a removable stylet

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, College of Medicine, Hanyang University, Seoul, Korea. jcshim@hanyang.ac.kr

Abstract

BACKGROUND
Electrical stimulation can be used to verify the location of epidural catheters. Although epiradicular localization of a catheter is important for patients with radiating pain in the lower extremities due to spinal stenosis or disc herniation, no prospective study results pertaining to the stimulation threshold within the epiradicular space have been reported. Therefore, we set out to investigate the threshold current for motor response in the epiradicular space by the use of a catheter containing a removable stylet. In addition, this study examines the reliability of such epiradicular stimulation.
METHODS
In 105 patients undergoing a retrograde interlaminar ventral epidural injection, a lumbar epiradicular catheter was inserted. Loss of resistance was used to identify the epidural space. Afterward, the TheraCath was advanced into the epiradicular space and connected to a nerve stimulator. A nerve stimulator delivered progressively, increasing electrical current until an appropriate muscle contraction was observed. Contrast media was then injected and epiradicular dispersal was assessed.
RESULTS
Epiradicular stimulation yielded a mean threshold of 0.56 +/- 0.50 mA (range: 0.06 to 2.84). When compared with epiradicular dispersal as a gold standard, the sensitivity, specificity, positive predictive value, and negative predictive value of epiradicular stimulation with the TheraCath were 90%, 56%, 96% and 33%, respectively.
CONCLUSIONS
The threshold current of an insulated needle required to elicit a motor response in the epiradicular space was evaluated. We conclude that the epiradicular stimulation test is useful for confirming catheter placement. Nonetheless, further studies are required before implementing its routine use in clinical settings.

Keyword

Epidural; Epiradicular; Reliability; Threshold current

MeSH Terms

Catheters
Contrast Media
Electric Stimulation
Epidural Space
Humans
Injections, Epidural
Lower Extremity
Muscle Contraction
Needles
Spinal Stenosis