Tuberc Respir Dis.  1995 Feb;42(1):84-92. 10.4046/trd.1995.42.1.84.

Nasal Continuous Positive Airway Pressure Titration and Time to Reach Optimal Pressure in Sleep Apnea Syndrome

Affiliations
  • 1Department of Internal Medicine, Yeungnam University, College of Medicine, Taegu, Korea.

Abstract

BACKGROUND
Nasal applied continuous positive airway pressure(CPAP) is a highly effective method of treatment for obstructive sleep apnea syndrome. More than a decade of accumulated experience with this treatment modality confirmed that it is unquestionably the medical treatment of choice for patients with obstructive sleep apnea syndrome. However it takes long time to reach optimal CPAP pressure. To save the time to reach optimal pressure, it is necessary to clarify the time to reach optimal pressure for treatment of obstructive sleep apnea syndrome. METHOD: CPAP pressure is titrated during an overnight study according to a standardized protocol. Just before the presleep bio-calibration procedures, the technician applies the nasal mask and switches on the clinical CPAP unit. Initial positive for pressure is typically 3.0 centimeters of water pressure. After sleep onset, the technician gradually increases the pressure until sleep- disordered breathing events disappear or become minimal. The pressure must maintain maximal airway patency during both NREM and REM sleep to be considered effective. Before recommending a final pressure setting, sleep recording and oximetry data are reviewed by an American Board of Sleep Medicine certified Sleep Specialist and a Registrered Polysomnographic Technologist.
RESULTS
We examined the time required to reach optimal pressure during routine CPAP titration in 127 consecutively evaluated individuals diagnosed with sleep-disordered breathing. Results indicate that 33% of patients required more than four hours to attain satisfactory titration. This indicates that a four-hour session is marginally enough time, at best, to determine a proper CPAP pressure setting. Moreover, 60 of 127 patients required further adjustment after optimal pressure was reached. These additional pressure trials were needed to confirm that higher pressures were not superior for eliminating sleep-disordered breathing events.
CONCLUSIONS
The data presented underscore the logistical difficulty of titrating CPAP during split-night studies without modifying the titration procedure. Futhermore, the time needed to reach optimal pressure makes it improbable that proper CPAP titration can be performed during a 2-3 hour nap study.

Keyword

Continuous positive airway pressure; Optimal pressure; Obstructive sleep apnea syndrome

MeSH Terms

Continuous Positive Airway Pressure*
Humans
Masks
Oximetry
Respiration
Sleep Apnea Syndromes*
Sleep Apnea, Obstructive
Sleep, REM
Specialization
Water
Water
Full Text Links
  • TRD
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr