End - tidal Carbon Dioxide Measurements in the Pediatric Patients : A Comparison of End - tidal PCO2 Sampled from the Proximal and Distal Ends of Pediatric Uncuffed Tracheal Tube

Affiliations
  • 1Department of Anesthesiology, College of Medicine, Korea University, Seoul, Korea.

Abstract

To determine the accuracy of end-tidal PCO2(PetCO2) measurement analysed with Ohmeda 5210 capnometer in the pediatric patients whose jung were ventilated with Ohio infant Circle Absorber and Ohio Ventilatior 7200, we compared PetCO2, measurements sampled from the proximal(PetCO2-p) and distal(PetCO2-d) ends of the tracheal tube to arterial PCO2(PaCO2) in 22 healthy pediatric patients between 6.8 to 18.5kg. The results were as follow: 1) Proximal PetCO2 and distal PetCO2 correlated with PaCO2(r=0.76%, SEE=2.79, r=0.70:SEE=3.01, respectively)(p<0.05). 2)Arterial to end-tidal PCO2 difference(delta(a-et)PCO2) was significantyly greater with distal(3.53+/-4.68 mmhg) than proximal(0.35+/-2.72 mmhg) sampling(p<0.05). 3) In subgroup of patients whose proximal to distal concentration of nitrous oxide difference was more than 2%, the delta(a-et)PCO2 using distal sampling(7.90+/-5.15 mmhg) was also significantly greater then it was using proximal sampling(-1.73+/-2.39 mmhg)(p<0.05).. We conclude that distal estimates of PetCO2-P can not provide accepatbale estimate of PaCO2 in healthy pediatric patients who are intubated with pediatric uncuffed tracheal tube.

Keyword

Anesthetic circuit; Ohio infant Circle Absorber; end-tidal PCO2 measurements; arterial PCO2

MeSH Terms

Carbon Dioxide*
Carbon*
Humans
Infant
Nitrous Oxide
Ohio
Carbon
Carbon Dioxide
Nitrous Oxide
Full Text Links
  • KJAE
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr